Determination of human blood glucose levels using microchip electrophoresis

Authors:

Abstract
A high‐performance monitoring system for human blood glucose levels was developed using microchip electrophoresis with a plastic chip. The combination of reductive amination as glucose labeling with fluorescent 2‐aminoacridone (AMAC) and glucose‐borate complex formation realized the highly selective detection of glucose even in a complex matrix such as a blood sample. The migration time of a single peak, observed on an electropherogram of AMAC‐labeled plasma, closely resembled that of glucose standard solution. The treatment of plasma with hexokinase or glucokinase for glucose phosphorylation resulted in a peak shift from approximately 145 to 70 s, corresponding to glucose and glucose‐6‐phosphate, respectively. A double‐logarithm plot revealed a linear relationship between glucose concentration and fluorescence intensity in the range of 1–300 μM of glucose (r2 = 0.9963; p p <0.05). The reproducibility of the assay was about 6.3–9.1% (RSD) and the within‐days and between‐days reproducibility were 1.6–8.4 and 5.2–7.2%, respectively. This system enables us to determine blood glucose with high sensitivity and accuracy, and will be applicable to clinical diagnosis.