Improved UV/O3 method for measuring the chemical oxygen demand
Authors:
- 2 January 2018
- journal article
- research article
- Published by IWA Publishing in Water Science & Technology
- Vol. 77 (5) , 1271-1279
Abstract
We improved the UV/O3-based method for measuring chemical oxygen demand (COD) in water. An on-line COD monitoring device was developed and the UV/O3 method was used to oxidize sample solutions. A model was established by using a support vector machines (SVM) algorithm to estimate dissolved oxygen and CO2 in solutions. Based on the measured data by each sensor during the oxidation process and the estimated dissolved oxygen and CO2, the UV/O3-based COD test accuracy was improved. This approach overcomes many problems associated with the conventional COD determination techniques such as long analysis time, consumption of expensive and toxic reagents, and production of secondary toxic waste. The effect of important parameters on the measurement of COD was systematically investigated. The Improved method was successfully applied to determine the COD of real samples from environmental water. Compared with the standard dichromate method, our UV/O3-based COD test method is more effective. The assay time of 10–15 min/sample can be readily achieved. A practical detection limit of 0.89 mg/L COD with a linear range of 1–300 mg/L was achieved under the normal conditions.Keywords
This publication has 18 references indexed in Scilit:
- Wastewater COD biodegradability fractionated by simple physical–chemical analysisChemical Engineering Journal, 2014
- The performance of electrochemical peroxidation process for COD reduction and biodegradability improvement of the wastewater from a paper recycling plantSeparation and Purification Technology, 2014
- Fenton treatment of landfill leachate under different COD loading factorsWaste Management, 2013
- Rapid and simple determination of chemical oxygen demand (COD) with potassium permanganate as an oxidizing agent by a flow-injection technique using a platinum-tube reactorBUNSEKI KAGAKU, 2004
- Dichromate method for the determination of chemical oxygen demandJournal of Analytical Chemistry, 2000
- Estimation of biochemical oxygen demand in slurry and effluents using ultra-violet spectrophotometryWater Research, 1997
- Kinetics of ozone photolysis in aqueous solutionAIChE Journal, 1996
- A rapid physical-chemical method for the determination of readily biodegradable soluble COD in municipal wastewaterWater Research, 1993
- Chemical oxygen demand (COD) using microwave digestionWater Research, 1989
- Dissolved organic carbon during a spring diatom bloom in Lake Mossø, DenmarkWater Research, 1982