Magnetic Resonance Perfusion Tracks 133 Xe Cerebral Blood Flow Changes After Carotid Stenting
- 1 March 2005
- journal article
- research article
- Published by Wolters Kluwer Health in Stroke
- Vol. 36 (3) , 676-678
Abstract
Background and Purpose— To compare magnetic resonance (MR) perfusion to gold-standard cerebral blood flow (CBF) determined by intra-arterial 133 Xe washout method. Methods— Eight patients with high-grade carotid stenoses underwent bolus-tracking MR perfusion and intra-arterial 133 Xe washout before and after carotid stenting. MR perfusion was compared with 133 Xe-CBF values using Pearson linear correlation analysis. Results— We observed a mean 37±38% increase in 133 Xe-CBF and a mean 19±27% increase in relative CBF (rCBF) by MR perfusion immediately after stent placement. Relative (percent) changes in MR-rCBF showed a close and linear correlation to those seen in 133 Xe-CBF ( r =0.91; R 2 =0.84; P =0.002). There was a trend for MR perfusion to underestimate change in CBF at higher relative changes in flow. Conclusion— Bolus-tracking MR perfusion correlates with 133 Xe-CBF in estimating postprocedural increases in blood flow but may underestimate the magnitude of the change with higher relative changes.Keywords
This publication has 5 references indexed in Scilit:
- Cerebral Hyperperfusion Syndrome after Percutaneous Transluminal Stenting of the Craniocervical ArteriesNeurosurgery, 2000
- Beneficial Effect of Carotid Endarterectomy in Symptomatic Patients with High-Grade Carotid StenosisNew England Journal of Medicine, 1991
- Intraoperative 133Xe Cerebral Blood Flow Measurements by Intravenous Versus Intracarotid MethodsAnesthesiology, 1990
- Cerebral blood flow: assessment with dynamic contrast-enhanced T2*-weighted MR imaging at 1.5 T.Radiology, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986