Reduction of fluoroscopic exposure for the air-contrast barium enema
Authors:
- Daniel R. BednarekDBDaniel R. Bednarek
- Stephen RudinSRStephen Rudin
- Roland WongRWRoland Wong
- Maria L. AndresMAMaria L. Andres
- 1 November 1983
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 56 (671) , 823-828
Abstract
In a fluoroscopic imaging system, image quality and patient dose are both affected by the optical system linking the image intensifier with the video camera. The effect on patient exposure of increasing the optical iris aperture size over that required for other procedures performed on the same imaging system was investigated for the air-contrast barium enema examination. Using a large-area transmission ionisation chamber to monitor the Roentgen-area-product of entrance exposure, a decrease in fluoroscopic radiation of greater than 50% was clinically documented for a fluoroscopic system utilising kVp and mA variable automatic brightness control. For this iris change, the video image was of acceptable quality for positioning and monitoring the patient, and no deleterious effect was detected in the conduct of the air-contrast exam. The availability of a variable-sized operator-selectable iris diaphragm would permit this dose-reduction approach to be extended to other fluoroscopic procedures.Keywords
This publication has 5 references indexed in Scilit:
- An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinationsPhysics in Medicine & Biology, 1982
- DETECTION OF CARCINOMA OF THE COLON AND RECTUM BY AIR CONTRAST ENEMA1981
- A Variable Aperture Fluoroscopic Unit for Reduced Patient ExposureRadiology, 1978
- The Absorption of X rays in a Body Equivalent PhantomBritish Journal of Radiology, 1965
- The Measurement of Patient DoseBritish Journal of Radiology, 1965