Computerized method for estimation of the location of a lung tumor on EPID cine images without implanted markers in stereotactic body radiotherapy
Authors:
- H ArimuraHAH Arimura
- Y EgashiraYEY Egashira
- Y ShioyamaYSY Shioyama
- K NakamuraKNK Nakamura
- S YoshidomeSYS Yoshidome
- S AnaiSAS Anai
- S NomotoSNS Nomoto
- H HondaHHH Honda
- F ToyofukuFTF Toyofuku
- Y HigashidaYHY Higashida
- Y OnizukaYOY Onizuka
- H TerashimaHTH Terashima
- 9 January 2009
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (3) , 665-677
Abstract
The purpose of this study was to develop a computerized method for estimation of the location of a lung tumor in cine images on an electronic portal imaging device (EPID) without implanted markers during stereotactic body radiotherapy (SBRT). Each tumor region was segmented in the first EPID cine image, i.e., reference portal image, based on a multiple-gray level thresholding technique and a region growing technique, and then the image including the tumor region was cropped as a 'tumor template' image. The tumor location was determined as the position in which the tumor template image took the maximum cross-correlation value within each consecutive portal image, which was acquired in cine mode on the EPID in treatment. EPID images with 512 x 384 pixels (pixel size: 0.56 mm) were acquired at a sampling rate of 0.5 frame s(-1) by using energies of 4, 6 or 10 MV on linear accelerators. We applied our proposed method to EPID cine images (226 frames) of 12 clinical cases (ages: 51-83, mean: 72) with a non-small cell lung cancer. As a result, the average location error between tumor points obtained by our method and the manual method was 1.47 +/- 0.60 mm. This preliminary study suggests that our method based on the tumor template matching technique might be feasible for tracking the location of a lung tumor without implanted markers in SBRT.Keywords
This publication has 16 references indexed in Scilit:
- Dosimetric impact of motion in free‐breathing and gated lung radiotherapy: A 4D Monte Carlo study of intrafraction and interfraction effectsMedical Physics, 2007
- Clinical Feasibility of Using an EPID in cine Mode for Image-Guided Verification of Stereotactic Body RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Automatic segmentation of phase‐correlated CT scans through nonrigid image registration using geometrically regularized free‐form deformationMedical Physics, 2007
- The management of respiratory motion in radiation oncology report of AAPM Task Group 76a)Medical Physics, 2006
- Design of 4D treatment planning target volumesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Tracking moving objects with megavoltage portal imaging: A feasibility studya)Medical Physics, 2006
- On the use of EPID‐based implanted marker tracking for 4D radiotherapyMedical Physics, 2004
- Computerized scheme for automated detection of lung nodules in low-dose computed tomography images for lung cancer screening1Academic Radiology, 2004
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002
- Development of a computerized method for identifying the posteroanterior and lateral views of chest radiographs by use of a template matching techniqueMedical Physics, 2002