Biotribological Behavior of Two Retrieved Implant Abutment Screws after Long-Term Use In Vivo
- 1 November 2012
- journal article
- research article
- Vol. 27 (6) , 1474-1480
Abstract
Purpose: To evaluate the effect of functional cyclic loading on two retrieved abutment screws used with single implant supporting cement-retained porcelain-fused-to-metal crowns by characterizing the structure, properties, and biotribologic behavior after long-term use in vivo. Materials and Methods: Two abutment screws were retrieved from the same patient and same implant at two different times. An external hex implant was used to replace a missing central incisor. A traumatic incident occurred 9 months after insertion and displaced the implant palatally by bending the screw. A second, similar incident occurred 31 months after insertion. In both cases, the abutment screws were retrieved and subjected to thorough nondestructive and destructive testing. Results: Light and scanning electron microscopic examinations revealed very minimal surface deterioration of the thread profile for the first screw (in service for 9 months) and demonstrated advanced metal adhesive wear in the form of galling for the second screw (in service for 31. months). The galling led to moderate thinning of the thread profile. Both screws were composed of Au-based alloy, where the microstructure of the matrix consisted of homogeneous equiaxed fine grains with two different second phase particles. Conclusion: It appears that the occurrence of adhesive wear on abutment screws in the form of galling is highly related to the length of in-service time in the mouth. This biotribologic behavior was inevitable and considered to be a normal consequence of long-term use in vivo. Metallurgic analysis indicated that both screws were identical in terms of composition and microstructure. INT J ORAL MAXILLOFAC IMPLANTS 2012;27:1474-1480Keywords
This publication has 9 references indexed in Scilit:
- Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implantsDental Materials Journal, 2009
- The Interface of Functional Biotribology and Regenerative Medicine in Synovial JointsTissue Engineering Part B: Reviews, 2008
- Influence of hardness on the wear resistance of 17-4 PH stainless steel evaluated by the pin-on-disc testingJournal of Materials Processing Technology, 2008
- A systematic review of the 5‐year survival and complication rates of implant‐supported single crownsClinical Oral Implants Research, 2007
- The effect of microstructure on the wear of cobalt-based alloys used in metal-on-metal hip implantsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2006
- Effect of varying load on wear resistance of carbon steel under unlubricated conditionsWear, 2003
- Galling initiation due to frictional heatingWear, 2003
- EXPERIMENTAL STUDY OF LOOSENING OF THREADED FASTENERS DUE TO DYNAMIC SHEAR LOADSJournal of Sound and Vibration, 2002
- Defects in hexed gold prosthetic screws: A metallographic and tensile analysisThe Journal of Prosthetic Dentistry, 2002