VERGLEICHENDE ANALYSE VON IN VITRO TESTVERFAHREN ZUR BEURTEILUNG DER HÄMOKOMPATIBILITÄT VON KARDIOVASKULÄREN STENTS
Authors:
- A. GebertAGA. Gebert
- D. BehrendDBD. Behrend
- G. ZinnerGZG. Zinner
- W. SchmidtWSW. Schmidt
- K.-P. SchmitzKSK.-P. Schmitz
- 1 January 2002
- journal article
- research article
- Published by Walter de Gruyter GmbH in Biomedical Engineering / Biomedizinische Technik
- Vol. 47 (s1b) , 827-830
Abstract
Within the scope of this study existing in vitro techniques for testing the hemocompatibility of coronary stents were analysed and optimised. Static and quasi-stationary systems were compared to a pulsed flow model with respect to platelet activity. The streamlines were visualized by dye injection. Blood flow was measured by ultrasonic Doppler velocity meter and electromagnetic flow meter. Uncoated stainless steel (316 L) stents were tested. Surrogate parameters of the hemocompatibility were the change in surface morphology after blood contact and the rise of biomechanical activation markers as C3a and beta-thromboglobulin. The results were correlated to the stent design and to the flow characteristics of the test systems.Keywords
This publication has 5 references indexed in Scilit:
- Multilink Stent Promotes Less Platelet and Leukocyte Adhesion than a Traditional Stainless Steel Stent: An in Vitro Experimental StudyJournal of Investigative Medicine, 2001
- Haemocompatibility of Endovascular Coronary Stents: Wiktor GX©. Hämokompatibilität von Koronarstents: Wiktor GX©Biomedical Engineering / Biomedizinische Technik, 2001
- Influence of Stent Length and Heparin Coating on Platelet ActivationThrombosis Research, 1999
- Hämokompatibilität verschiedener Kunststoffgranulate - Haemocompatibility of Various Polymer GranulatesBiomedical Engineering / Biomedizinische Technik, 1997
- In vitro model to test the thrombogenicity of coronary stentsThrombosis Research, 1994