Analytical and clinical performance of a homogeneous enzymatic LDL-cholesterol assay compared with the ultracentrifugation-dextran sulfate-Mg2+ method
Authors:
Open Access
- 1 June 1998
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 44 (6) , 1242-1250
Abstract
LDL-cholesterol (LDL-C) concentration is currently determined in most clinical laboratories by the Friedewald calculation. This approach has several limitations and may not meet the current total error requirement in LDL-C measurement of ≤ 12%. We evaluated the analytical and clinical performance of the direct N-geneous LDL-C assay (Equal Diagnostics). The N-geneous method correlated highly with the modified beta-quantification assay (r = 0.95; y = 0.91x + 70.6 mg/L; n = 199), showed no significant effect of increased triglyceride or other common interferants, and performed adequately in serum samples from nonfasting individuals. This assay demonstrated a mean total error of 6.75% over a wide range of LDL-C concentrations. In addition, at the medical decision cutoff points, this LDL-C assay showed positive predictive values of 78–95% and negative predictive values of 84–99%. We conclude that the N-geneous LDL-C meets the currently established analytical performance goals and appears to have a role in the diagnosis and management of hypercholesterolemic patients.Keywords
This publication has 11 references indexed in Scilit:
- Evaluation and Clinical Application of a Direct Low-Density Lipoprotein Cholesterol Assay in Normolipidemic and Hyperlipidemic AdultsThe American Journal of Cardiology, 1997
- Analytical performance and clinical utility of a direct LDL-cholesterol assay in a hyperlipidemic pediatric populationClinical Chemistry, 1996
- Summary of the Second Report of the National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel II)JAMA, 1993
- Ascorbic acid and dehydroascorbic acid analyses in biological samplesAnalytical Biochemistry, 1992
- Measurement of low-density-lipoprotein cholesterol in serum: a status reportClinical Chemistry, 1992
- Lipoprotein cholesterol concentrations in the plasma of human subjects as measured in the fed and fasted states.Clinical Chemistry, 1988
- Evaluation of a commercial reagent for precipitating human serum low-density lipoproteinClinica Chimica Acta, 1985
- Five methods for determining low-density lipoprotein cholesterol compared.Clinical Chemistry, 1984
- An evaluation of three commercial methods for the determination of LDL-cholesterolClinica Chimica Acta, 1984
- Spectrophotometry of Proline in Plasma and UrineClinical Chemistry, 1972