The within‐ and between‐laboratories reproducibility and predictive capacity of Amino acid Derivative Reactivity Assay using 4 mM test chemical solution: Results of ring study implemented at five participating laboratories
- 1 December 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Toxicology
- Vol. 42 (2) , 318-333
Abstract
Amino acid derivative reactivity assay (ADRA) for skin sensitization was adopted as an alternative method in the 2019 OECD Guideline for the Testing of Chemicals (OECD TG 442C). The molar ratio of the nucleophilic reagent to the test chemicals in the reaction solution was set to 1:50. Imamura et al. reported that changing this molar ratio from 1:50 to 1:200 reduced in false negatives and improved prediction accuracy. Hence, a ring study using ADRA with 4 mM of a test chemical solution (ADRA, 4 mM) was conducted at five different laboratories to verify within‐ and between‐laboratory reproducibilities (WLR and BLR, respectively). In this study, we investigated the WLR and BLR using 14 test chemicals grouped into three classes: (1) eight proficiency substances, (2) four test chemicals that showed false negatives in the ADRA with 1 mM test chemical solution (ADRA, 1 mM), but correctly positive in ADRA (4 mM), and (3) current positive control (phenylacetaldehyde) and a new additional positive control (squaric acid diethyl ester). The results showed 100% reproducibility and 100% accuracy for skin sensitization. Hence, it is clear that the ADRA (4 mM) is an excellent test method in contrast to the currently used ADRA (1 mM). We plan to resubmit the ADRA (4 mM) test method to the OECD Test Guideline Group in the near future so that OECD TG 442C could be revised for the convenience and benefit of many ADRA users.Keywords
This publication has 13 references indexed in Scilit:
- Precipitation of test chemicals in reaction solutions used in the amino acid derivative reactivity assay and the direct peptide reactivity assayJournal of Pharmacological and Toxicological Methods, 2019
- A newly developed means of HPLC-fluorescence analysis for predicting the skin sensitization potential of multi-constituent substances using ADRAToxicology in Vitro, 2019
- The within‐ and between‐laboratory reproducibility and predictive capacity of the in chemico amino acid derivative reactivity assay: Results of validation study implemented in four participating laboratoriesJournal of Applied Toxicology, 2019
- Expanding the applicability of the amino acid derivative reactivity assay: Determining a weight for preparation of test chemical solutions that yield a predictive capacity identical to the conventional method using molar concentration and demonstrating the capacity to detect sensitizers in liquid mixturesJournal of Pharmacological and Toxicological Methods, 2019
- The underlying factors that explain why nucleophilic reagents rarely co-elute with test chemicals in the ADRAJournal of Pharmacological and Toxicological Methods, 2019
- Cause of and countermeasures for oxidation of the cysteine‐derived reagent used in the amino acid derivative reactivity assayJournal of Applied Toxicology, 2018
- Measurement and Modeling of Extra-Column Effects Due to Injection and Connections in Capillary Liquid ChromatographySeparations, 2015
- A novel in chemico method to detect skin sensitizers in highly diluted reaction conditionsJournal of Applied Toxicology, 2015
- Development of a prediction method for skin sensitization using novel cysteine and lysine derivativesJournal of Pharmacological and Toxicological Methods, 2014
- Investigation of the Kinetics and Mechanism of Acid Chloride Hydrolysis in an Oil/Water System Using Microelectrochemical Measurements at Expanding Droplets (MEMED)Langmuir, 2004