Technical Report: Reference photon dosimetry data for Varian accelerators based on IROC-Houston site visit data
- 15 April 2016
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 43 (5) , 2374-2386
Abstract
Purpose: Accurate data regarding linear accelerator(Linac) radiation characteristics are important for treatment planning system modeling as well as regular quality assurance of the machine. The Imaging and Radiation Oncology Core-Houston (IROC-H) has measured the dosimetric characteristics of numerous machines through their on-site dosimetry review protocols. Photon data are presented and can be used as a secondary check of acquired values, as a means to verify commissioning a new machine, or in preparation for an IROC-H site visit. Methods: Photon data from IROC-H on-site reviews from 2000 to 2014 were compiled and analyzed. Specifically, data from approximately 500 Varian machines were analyzed. Each dataset consisted of point measurements of several dosimetric parameters at various locations in a water phantom to assess the percentage depth dose, jaw output factors, multileaf collimator small field output factors, off-axis factors, and wedge factors. The data were analyzed by energy and parameter, with similarly performing machine models being assimilated into classes. Common statistical metrics are presented for each machine class. Measurement data were compared against other reference data where applicable. Results: Distributions of the parameter data were shown to be robust and derive from a student’s t distribution. Based on statistical and clinical criteria, all machine models were able to be classified into two or three classes for each energy, except for 6 MV for which there were eight classes. Quantitative analysis of the measurements for 6, 10, 15, and 18 MV photon beams is presented for each parameter; supplementary material has also been made available which contains further statistical information. Conclusions: IROC-H has collected numerous data on Varian Linacs and the results of photon measurements from the past 15 years are presented. The data can be used as a comparison check of a physicist’s acquired values. Acquired values that are well outside the expected distribution should be verified by the physicist to identify whether the measurements are valid. Comparison of values to this reference data provides a redundant check to help prevent gross dosimetric treatment errors.Keywords
Funding Information
- National Cancer Institute (NCI) (CA180803)
- National Cancer Institute (CA180803)
- National Cancer Institute (CA180803)
This publication has 14 references indexed in Scilit:
- Erratum: “The Radiological Physics Center's standard dataset for small field size output factors”Journal of Applied Clinical Medical Physics, 2014
- Commissioning of the Varian TrueBeam linear accelerator: A multi‐institutional studyMedical Physics, 2013
- Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear acceleratorsJournal of Applied Clinical Medical Physics, 2013
- Commissioning and dosimetric characteristics of TrueBeam system: Composite data of three TrueBeam machinesMedical Physics, 2012
- The Radiological Physics Center's standard dataset for small field size output factorsJournal of Applied Clinical Medical Physics, 2012
- A beam-matching concept for medical linear acceleratorsActa Oncologica, 2009
- Comparative measurements on a series of accelerators by the same vendorMedical Physics, 1999
- Beam characteristics of a new model of 6-MV linear acceleratora)Medical Physics, 1992
- Dosimetry of a dual photon energy linear acceleratorMedical Physics, 1987
- Characteristics of the 6-MV photon beam produced by a dual energy linear acceleratorMedical Physics, 1987