Integrated Sample Cleanup−Capillary Electrophoresis Microchip for High-Performance Short Tandem Repeat Genetic Analysis
Authors:
- 5 December 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 81 (1) , 210-217
Abstract
An integrated PCR sample cleanup and preconcentration process is developed for forensic short tandem repeat (STR) analysis using a streptavidin-modified photopolymerized capture gel injector for microchip capillary electrophoresis (μCE). PCR samples generated with one biotinylated primer and one fluorescent primer provide the input to the streptavidin-based affinity capture-μCE device. Monoplex PCR samples processed by the device exhibited ∼10- to 50-fold increased fluorescence intensities, and DNA profiles generated using 9-plex STR samples displayed ∼14- to 19-fold higher signal intensities compared to those analyzed using traditional cross injection. Complete STR profiles were obtained with as few as 25 copies of DNA template using the capture-μCE device. Four DNA samples with various degrees of degradation were also tested. Samples analyzed using the capture-μCE device resulted in a significant increase of successful allele detection. The ability of our capture-μCE device and method to remove contaminating ions, to concentrate the sample injection plug, and to eliminate electrokinetic injection bias provides a powerful approach for integrating sample cleanup with DNA separation.Keywords
This publication has 32 references indexed in Scilit:
- Real-time forensic DNA analysis at a crime scene using a portable microchip analyzerForensic Science International: Genetics, 2008
- A Forensic Laboratory Tests the Berkeley Microfabricated Capillary Array Electrophoresis Device*Journal of Forensic Sciences, 2008
- Forensic DNA Analysis on Microfluidic Devices: A ReviewJournal of Forensic Sciences, 2007
- Integrated Portable Polymerase Chain Reaction-Capillary Electrophoresis Microsystem for Rapid Forensic Short Tandem Repeat TypingAnalytical Chemistry, 2007
- A fully integrated microfluidic genetic analysis system with sample-in–answer-out capabilityProceedings of the National Academy of Sciences, 2006
- Rapid and High‐Throughput Forensic Short Tandem Repeat Typing Using a 96‐Lane Microfabricated Capillary Array Electrophoresis Microdevice*Journal of Forensic Sciences, 2006
- Analysis of Sources of Error in Quantitation of Purified DNA Fragments and Unpurified PCR Products by DNA Microchip ElectrophoresisChromatographia, 2005
- A high‐performance multilane microdevice system designed for the DNA forensics laboratoryElectrophoresis, 2004
- Micromachining a Miniaturized Capillary Electrophoresis-Based Chemical Analysis System on a ChipScience, 1993
- Bias in quantitative capillary zone electrophoresis caused by electrokinetic sample injectionAnalytical Chemistry, 1988