The ultimate signal‐to‐noise ratio in realistic body models
Authors:
- 4 December 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 78 (5) , 1969-1980
Abstract
Purpose: We compute the ultimate signal‐to‐noise ratio (uSNR) and G‐factor (uGF) in a realistic head model from 0.5 to 21 Tesla.Methods: We excite the head model and a uniform sphere with a large number of electric and magnetic dipoles placed at 3 cm from the object. The resulting electromagnetic fields are computed using an ultrafast volume integral solver, which are used as basis functions for the uSNR and uGF computations.Results: Our generalized uSNR calculation shows good convergence in the sphere and the head and is in close agreement with the dyadic Green's function approach in the uniform sphere. In both models, the uSNR versus B0trend was linear at shallow depths and supralinear at deeper locations. At equivalent positions, the rate of increase of the uSNR with B0was greater in the sphere than in the head model. The uGFs were lower in the realistic head than in the sphere for acceleration in the anterior‐posterior direction, but similar for the left‐right direction.Conclusion: The uSNR and uGFs are computable in nonuniform body models and provide fundamental performance limits for human imaging with close‐fitting MRI array coils. Magn Reson Med 78:1969–1980, 2017. © 2016 International Society for Magnetic Resonance in Medicine.Keywords
Funding Information
- National Institutes of Health (R01EB006847, P41EB015896 and K99EB019482)
This publication has 42 references indexed in Scilit:
- Size‐optimized 32‐channel brain arrays for 3 T pediatric imagingMagnetic Resonance in Medicine, 2011
- Performance evaluation of a 32‐element head array with respect to the ultimate intrinsic SNRNMR in Biomedicine, 2009
- Electrodynamic constraints on homogeneity and radiofrequency power deposition in multiple coil excitationsMagnetic Resonance in Medicine, 2009
- Parallel RF transmission with eight channels at 3 TeslaMagnetic Resonance in Medicine, 2006
- 32‐element receiver‐coil array for cardiac imagingMagnetic Resonance in Medicine, 2006
- Design considerations and coil comparisons for 7 T brain imagingApplied Magnetic Resonance, 2005
- Analysis of wave behavior in lossy dielectric samples at high fieldMagnetic Resonance in Medicine, 2002
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Specific absorption rates and induced current densities for an anatomy-based model of the human for exposure to time-varying magnetic fields of MRIMagnetic Resonance in Medicine, 1999
- Calculation of the signal-to-noise ratio for simple surface coils and arrays of coils [magnetic resonance imaging]IEEE Transactions on Biomedical Engineering, 1995