Comparison of KAP values for CAG examinations: the influence of parameter settings
Author:
- 16 February 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 139 (1-3) , 363-366
Abstract
This study shows that changing settings like detector dose, added copper filtration, kV–mA–ms settings and frame rate can substantially reduce the average air kerma-area product value per examination in coronary angiography. The default factory settings of cardiac X-ray systems are often not optimised and therefore the dose settings are mostly too high. The type of modifications that are allowed depends on the X-ray system. With the GE Innovation2000, the freedom of making adjustments is restricted. With the Siemens Coroscop and the Philips Allura Xper FD10, several adjustments can be made. It is possible to lower the dose to patient and staff by not unreservedly accept the default factory settings. However, the point of discussion is ‘how low should you go’, because the needed clinical image quality is difficult to define objectively using phantom measurements.Keywords
This publication has 4 references indexed in Scilit:
- A large-scale multicentre study in Belgium of dose area product values and effective doses in interventional cardiology using contemporary X-ray equipmentRadiation Protection Dosimetry, 2007
- Factors That Influence Radiation Dose in Percutaneous Coronary InterventionJournal of Interventional Cardiology, 2006
- Is patient size important in dose determination and optimization in cardiology?Physics in Medicine & Biology, 2003
- Variability of radiation output dynamic range in modern cardiac catheterization imaging systemsCatheterization and Cardiovascular Diagnosis, 1998