Maltose Interference-Free Test Strips for Blood Glucose Testing at Point-of-Care: A Laboratory Performance Evaluation
Authors:
- 1 November 2010
- journal article
- research article
- Published by SAGE Publications in Diabetes Technology & Therapeutics
- Vol. 12 (11) , 889-893
Abstract
Background: Maltose interference is a concern with blood glucose testing at point-of-care. We evaluated a maltose interference-free test strip (with a modified recombinant glucose dehydrogenase-pyrroloquinoline quinone system) for the Accu-Chek ® Performa glucose meter (Roche Diagnostics, Mannheim, Germany). Methods: Blood specimens ( n = 120) sent for clinical laboratory glucose testing were used in assessing performance characteristics, including imprecision, linearity, clinical impact analysis, and method comparison, of the test strips. To evaluate sugar interference, two heparinized blood specimens were spiked with maltose, xylose, and galactose (up to 500 mg/dL) followed by testing with modified Performa, Accutrend ® (Roche Diagnostics), and Advantage II (Roche Diagnostics) test strips and by the laboratory method. Results: Test strips demonstrated total laboratory coefficients of variation of 20% from the laboratory result. Conclusions: The modified and improved Performa test strips were not affected by maltose and xylose. They meet ISO 15197:2003 requirements with a slight bias (−0.5 mmol/L) compared to the laboratory method.Keywords
This publication has 8 references indexed in Scilit:
- Comparison of blood glucose meters in a New Zealand diabetes centreAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2009
- Interference of Maltose, Icodextrin, Galactose, or Xylose with Some Blood Glucose Monitoring SystemsPharmacotherapy: The Journal of Human Pharmacology and Drug Therapy, 2007
- Hypoglycemic Coma in a Diabetic Patient on Peritoneal Dialysis due to Interference of Icodextrin Metabolites With Capillary Blood Glucose MeasurementsDiabetes Care, 2004
- Pharmacokinetics of icodextrin in peritoneal dialysis patientsKidney International, 2002
- General Deming Regression for Estimating Systematic Bias and Its Confidence Interval in Method-Comparison StudiesClinical Chemistry, 2000
- Icodextrin use in CCPD patients during peritonitis: ultrafiltration and serum disaccharide concentrationsNephrology Dialysis Transplantation, 1998
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986