A Randomized Prospective Study of Single Coil Versus Dual Coil Defibrillation in Patients with Ventricular Arrhythmias Undergoing Implantable Cardioverter Defibrillator Therapy
Authors:
- 24 July 2003
- journal article
- clinical trial
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 26 (8) , 1684-1690
Abstract
ICD implantation is standard therapy for malignant ventricular arrhythmias. The advantage of dual and single coil defibrillator leads in the successful conversion of arrhythmias is unclear. This study compared the effectiveness of dual versus single coil defibrillation leads. The study was a prospective, multicenter, randomized study comparing a dual with a single coil defibrillation system as part of an ICD using an active pectoral electrode. Seventy‐six patients (64 men, 12 women; age 61 ± 11 years ) were implanted with a dual (group 1, n = 38) or single coil lead system (group 2, n = 38 ). The patients represented a typical ICD cohort: 60% presented with ischemic cardiomyopathy as their primary cardiac disease, the mean left ventricular ejection fraction was 0.406 ± 0.158 . The primary tachyarrhythmia was monomorphic ventricular tachyarrhythmia in 52.6% patients and ventricular fibrillation in 38.4%. There was no significant difference in terms of P and R wave amplitudes, pacing thresholds, and lead impedance at implantation and follow‐up in the two groups. There was similarly no difference in terms of defibrillation thresholds (DFT) at implantation. Patients in group 1 had an average DFT of 10.2 ± 5.2 J compared to 10.3 ± 4.1 J in Group 2, P = NS. This study demonstrates no significant advantage of a dual coil lead system over a single coil system in terms of lead values and defibrillation thresholds. This may have important bearing on the choice of lead systems when implanting ICDs. (PACE 2003; 26:1684–1690)Keywords
This publication has 17 references indexed in Scilit:
- Use of a Single Coil Transvenous Electrode with an Abdominally Placed Implantable Cardioverter Defibrillator in ChildrenPacing and Clinical Electrophysiology, 2000
- A Randomized Study of the Prevention of Sudden Death in Patients with Coronary Artery DiseaseNew England Journal of Medicine, 1999
- Atrial defibrillation with a transvenous leadJACC, 1999
- Defibrillation Efficacy of Different Electrode Placements in a Human Thorax ModelPacing and Clinical Electrophysiology, 1999
- Improved Survival with an Implanted Defibrillator in Patients with Coronary Disease at High Risk for Ventricular ArrhythmiaNew England Journal of Medicine, 1996
- Transvenous Defibrillation Lead SystemsJournal of Cardiovascular Electrophysiology, 1996
- Efficacy of a single-lead unipolar transvenous defibrillator compared with a system employing an additional coronary sinus electrode. A prospective, randomized study.Circulation, 1994
- Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode.Circulation, 1994
- A simplified, single-lead unipolar transvenous cardioversion-defibrillation system.Circulation, 1993
- Epicardial mapping of ventricular defibrillation with monophasic and biphasic shocks in dogs.Circulation Research, 1993