The Effect of Instantaneous Input Dynamic Range Setting on the Speech Perception of Children with the Nucleus 24 Implant
- 1 June 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Ear & Hearing
- Vol. 30 (3) , 340-349
Abstract
Objective: The purpose of this study was to examine the effects of a wider instantaneous input dynamic range (IIDR) setting on speech perception and comfort in quiet and noise for children wearing the Nucleus 24™ implant system and the Freedom™ speech processor. In addition, children's ability to understand soft and conversational level speech in relation to aided sound-field thresholds was examined. Design: Thirty children (age, 7 to 17 years) with the Nucleus 24 cochlear implant system and the Freedom speech processor with two different IIDR settings (30 versus 40 dB) were tested on the Consonant Nucleus Consonant (CNC) word test at 50 and 60 dB SPL, the Bamford-Kowal-Bench Speech in Noise Test, and a loudness rating task for four-talker speech noise. Aided thresholds for frequency-modulated tones, narrowband noise, and recorded Ling sounds were obtained with the two IIDRs and examined in relation to CNC scores at 50 dB SPL. Speech Intelligibility Indices were calculated using the long-term average speech spectrum of the CNC words at 50 dB SPL measured at each test site and aided thresholds. Results: Group mean CNC scores at 50 dB SPL with the 40 IIDR were significantly higher (p < 0.001) than with the 30 IIDR. Group mean CNC scores at 60 dB SPL, loudness ratings, and the signal to noise ratios-50 for Bamford-Kowal-Bench Speech in Noise Test were not significantly different for the two IIDRs. Significantly improved aided thresholds at 250 to 6000 Hz as well as higher Speech Intelligibility Indices afforded improved audibility for speech presented at soft levels (50 dB SPL). Conclusion: These results indicate that an increased IIDR provides improved word recognition for soft levels of speech without compromising comfort of higher levels of speech sounds or sentence recognition in noise.Keywords
This publication has 26 references indexed in Scilit:
- Clinical Evaluation of Expanded Input Dynamic Range in Nucleus Cochlear ImplantsEar & Hearing, 2007
- Recognition of Speech Presented at Soft to Loud Levels by Adult Cochlear Implant Recipients of Three Cochlear Implant SystemsEar & Hearing, 2004
- Optimizing Dynamic Range in Children Using the Nucleus Cochlear ImplantEar & Hearing, 2004
- Effects of Presentation Level on Phoneme and Sentence Recognition in Quiet by Cochlear Implant ListenersEar & Hearing, 2003
- Methods and Applications of the Audibility Index in Hearing Aid Selection and FittingTrends in Amplification, 2002
- Effects of the Acoustical Dynamic Range on Speech Recognition with Cochlear Implants:Efectos en el rango dinámico del reconocimiento del habla con implantes coclearesInternational Journal of Audiology, 2001
- Children with Minimal Sensorineural Hearing Loss: Prevalence, Educational Performance, and Functional StatusEar & Hearing, 1998
- Performance of children aged 9 to 17 years on a test of speech intelligibility in noise using sentence material with controlled word predictabilityThe Journal of the Acoustical Society of America, 1979
- Preferred Method For Clinical Determination Of Pure-Tone ThresholdsJournal of Speech and Hearing Disorders, 1959
- Factors Governing the Intelligibility of Speech SoundsThe Journal of the Acoustical Society of America, 1947