Air-Puff-Induced Dynamics of Ocular Components Measured with Optical Biometry
Open Access
- 1 May 2019
- journal article
- research article
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Investigative Ophthalmology & Visual Science
- Vol. 60 (6) , 1979-1986
Abstract
Purpose: To analyze the dynamics of all optical components of the eye and the behavior of the eyeball under air-puff conditions in vivo. To determine the impact of the intraocular pressure (IOP) on the air-puff-induced deformation of the eye. Methods: Twenty eyes of 20 healthy subjects were included in this study. The dynamics of the ocular components, such as the cornea, the crystalline lens, and the retina, was measured by a prototype swept source optical coherence tomography biometer integrated with the air-puff system. The system allows to acquire a series of axial scans at the same location as a function of time with no transverse scanning. Several parameters were extracted from optical coherence tomography data. The IOP was measured using a Goldmann applanation tonometry. The measurements of the eyes were performed before and 2 hours after administration of IOP-reducing drops, namely, 0.2 % brimonidine tartrate. Results: There is a statistically significant correlation of corneal thickness, vitreous depth, and eye length with IOP. The deformation amplitudes of the cornea and the crystalline lens are inversely proportional to the IOP, but statistical significance is achieved only for the cornea. The crystalline lens is displaced without compression, and the return has the form of wobbling. The reduction of IOP level induces corresponding changes in the extracted parameters. Conclusions: Optical biometry combined with air puff provides comprehensive information on the in vivo behavior of all ocular components, including the crystalline lens. Measurement of the axial length dynamics of during deformation enables correcting the deformation for eye retraction.Keywords
This publication has 40 references indexed in Scilit:
- Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjectsBMC Ophthalmology, 2017
- Quantitative assessment of responses of the eyeball based on data from the Corvis tonometerComputers in Biology and Medicine, 2015
- Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucomaBritish Journal of Ophthalmology, 2013
- A New Tonometer—The Corvis ST Tonometer: Clinical Comparison with Noncontact and Goldmann Applanation TonometersInvestigative Ophthalmology & Visual Science, 2013
- A History of Intraocular Pressure and Its MeasurementOptometry and Vision Science, 2011
- Intraocular Pressure Measurement Precision with the Goldmann Applanation, Dynamic Contour, and Ocular Response Analyzer TonometersOphthalmology, 2010
- The use of the Reichert ocular response analyser to establish the relationship between ocular hysteresis, corneal resistance factor and central corneal thickness in normal eyesContact Lens and Anterior Eye, 2006
- Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation TonometryArchives of Ophthalmology (1950), 2006
- Determining in vivo biomechanical properties of the cornea with an ocular response analyzerJournal of Cataract & Refractive Surgery, 2005
- The Effect of Corneal Thickness on Applanation TonometryAmerican Journal of Ophthalmology, 1993