Comparison of Age-derived Lens Thickness to Optically Measured Lens Thickness in IOL Power Calculation: A Clinical Study
Author:
- 1 February 2012
- journal article
- research article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 28 (2) , 154-155
Abstract
To assess whether lens thickness derived from age should be used in intraocular lens (IOL) power calculation. Ninety-three eyes underwent optical low coherence reflectometry (OLCR) prior to cataract surgery. Lens thickness (LT) by OLCR was used to calculate the IOL power. The mean absolute refractive error (MAE) and mean refractive error (ME) were calculated. Regression analysis between age and LT was performed. The expected IOL power, MAE, and ME were calculated with two age-derived lens thickness formulae: LT=4+(age/100) and LT=3.81+(0.0119×age). Lens thickness by OLCR was not significantly different from that by age, but the IOL power was different. Refractive outcome was better with OLCR-based LT. Correlation between age and LT was not as linear as implied by the formulae. In its current form, age-derived LT should not be used in IOL power calculation.Keywords
This publication has 9 references indexed in Scilit:
- Refractive outcomes after cataract surgery: Scheimpflug keratometry versus standard automated keratometry in virgin corneasJournal of Cataract & Refractive Surgery, 2011
- Comparison of Two Partial Coherence Interferometry Devices for Ocular BiometryKlinische Monatsblätter für Augenheilkunde, 2010
- Comparison of 2 laser instruments for measuring axial lengthJournal of Cataract & Refractive Surgery, 2010
- Intraocular lens power calculation: Clinical comparison of 2 optical biometry devicesJournal of Cataract & Refractive Surgery, 2010
- Comparison and Evaluation of Ocular Biometry Using a New Noncontact Optical Low-Coherence ReflectometerOphthalmology, 2009
- A new optical low coherence reflectometry device for ocular biometry in cataract patientsBritish Journal of Ophthalmology, 2009
- Accuracy of a new partial coherence interferometry analyser for biometric measurementsBritish Journal of Ophthalmology, 2009
- Prediction of the effective postoperative (intraocular lens) anterior chamber depthJournal of Cataract & Refractive Surgery, 2006
- Axial Dimension of the Human Cataractous LensArchives of Ophthalmology (1950), 1993