Sample transport by pneumatic tube system alters results of multiple electrode aggregometry but not rotational thromboelastometry
Authors:
- 29 August 2012
- journal article
- Published by Taylor & Francis in Platelets
- Vol. 24 (6) , 454-461
Abstract
Pneumatic tube systems (PTS) present a convenient way for blood sample transport in medical facilities. Associated preanalytical interference in various tests is largely unknown. Implementing point-of-care coagulation management at our institution, we investigated multiple electrode aggregometry (MEA) and rotational thromboelastometry (ROTEM) after PTS transportation. Whole blood samples from patients undergoing general or trauma surgery were analysed by MEA after collection (baseline, ‘0 × PTS’) and sent on a predefined PTS track (n = 12). MEA was repeated after samples travelled the track 4 (‘4 × PTS’), 8 (‘8 × PTS’) and 12 times (‘12 × PTS’) and compared with stationary controls analysed at the same time. Samples for ROTEM (n = 6) were analysed after collection and travelling the track 12 times. An acceleration detector recorded g-forces on the PTS track. At ‘0 × PTS’ no significant differences in MEA results were detected. Values were significantly lower for transported samples compared with controls (‘4 × PTS’ to ‘12 × PTS’, p < 0.001). Furthermore, MEA results of PTS samples were significantly decreased for ‘4 × PTS’ to ‘12 × PTS’ compared to baseline (p < 0.001). Except for the clotting time in EXTEM PTS transport did not significantly alter results for investigated ROTEM parameters, compared with baseline and stationary controls. Acceleration detector readout revealed alternating g-forces between −6.3 and +5.9 during transport. PTS transport caused invalid results in MEA while only one ROTEM parameter was found to be affected in this study. Variable acceleration during transport provides a potential reason for platelet activation. The authors recommend sample transport by hand or the device to be placed patient-side when MEA is performed.Keywords
This publication has 32 references indexed in Scilit:
- First-line Therapy with Coagulation Factor Concentrates Combined with Point-of-Care Coagulation Testing Is Associated with Decreased Allogeneic Blood Transfusion in Cardiovascular SurgeryAnesthesiology, 2011
- Laboratory-based ROTEM® analysis: implementing pneumatic tube transport and real-time graphic transmissionInternational Journal of Laboratory Hematology, 2011
- Transfusion in trauma: thromboelastometry-guided coagulation factor concentrate-based therapy versus standard fresh frozen plasma-based therapyCritical Care, 2011
- In Vitro impairment of whole blood coagulation and platelet function by hypertonic saline hydroxyethyl starchScandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 2011
- Goal-directed coagulation management of major trauma patients using thromboelastometry (ROTEM®)-guided administration of fibrinogen concentrate and prothrombin complex concentrateCritical Care, 2010
- Assessment of Platelet Function in Whole Blood by Multiple Electrode Aggregometry: Transport of Samples Using a Pneumatic Tube SystemAmerican Journal of Clinical Pathology, 2009
- Platelet concentrates transfusion in cardiac surgery and platelet function assessment by multiple electrode aggregometryActa Anaesthesiologica Scandinavica, 2009
- An Evaluation of Cyclooxygenase-1 Inhibition Before Coronary Artery Surgery: Aggregometry Versus Patient Self-ReportingAnesthesia & Analgesia, 2008
- Utility of the PFA-100 Instrument and the Novel Multiplate Analyzer for the Assessment of Aspirin and Clopidogrel Effects on Platelet Function in Patients With Cardiovascular DiseaseClinical and Applied Thrombosis/Hemostasis, 2008
- Blutgerinnungsstudien mit der Thrombelastographie, einem neuen UntersuchungsverfahrenJournal of Molecular Medicine, 1948