The Effect of Serum Protein Concentration on Wear Rates in a Hip Simulator
- 15 October 2009
- journal article
- Published by SAGE Publications in Journal of Biomaterials Applications
- Vol. 25 (2) , 145-159
Abstract
Ultra high molecular weight polyethylene cups, 22 mm in diameter, were aged for 5 years in the normal laboratory environment. Half of the samples had been processed by the standard radiation sterilization techniques, while the remainder had been cross-linked by a technique involving higher radiation doses and controlled temperature at the time of irradiation. The samples had been tested in a hip simulator for 5,000,000 cycles using a lubricant that had been diluted 1:1 with deionized water. Once that testing was completed, further testing was conducted using lubricant with greater and lesser serum protein concentrations, and the results compared with those that had already been recorded. Comparison of the wear rates within the study as well as to published data concerning the effect of serum concentration showed results that were consistent with assumed differences in lubrication ability at different concentrations. The results of other published studies were found to be inconsistent with each other and different from some of the results of this study. There is shown to be a need for carefully controlled and conducted studies to agree, if possible, on the importance of the serum concentration and the appropriate parameters to be used in testing, as well as variations that may be necessary with different bearing material characteristics.Keywords
This publication has 15 references indexed in Scilit:
- A Review of Lubrication Conditions for Wear Simulation in Artificial Hip ReplacementsTribology Transactions, 2006
- Effect of low protein concentration lubricants in hip simulatorsJournal of Orthopaedic Science, 2006
- Characterization of the Wear Performance of Crosslinked UHMWPE and Relationship to Molding Procedures2004
- Effect of Lubricant Protein Concentration on the Wear of Ultra-High Molecular Weight Polyethylene Sliding Against a CoCr CounterfaceJournal of Tribology, 2003
- Wear of ultra-high molecular weight polyethylene and polytetrafluoroethylene in a hip simulator: A dose-response study of protein concentrationActa Orthopaedica, 2000
- Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra high molecular weight polyethylene acetabular cups in a hip joint simulatorJournal of Orthopaedic Research, 1999
- Effect of protein lubrication on the wear properties of materials for prosthetic jointsJournal of Biomedical Materials Research, 1999
- Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditionsBiomaterials, 1996
- Low wear rate of UHMWPE against zirconia ceramic (Y‐PSZ) in comparison to alumina ceramic and SUS 316L alloyJournal of Biomedical Materials Research, 1991
- Wear characteristics of UHMW polyethylene: A method for accurately measuring extremely low wear ratesJournal of Biomedical Materials Research, 1978