Cartilage Thickness: Factors Influencing Multidetector CT Measurements in a Phantom Study

Authors:

Abstract
Purpose: To prospectively assess in a phantom the reconstruction errors and detection limits of cartilage thickness measurements obtained with multidetector computed tomographic (CT) arthrography, as a function of contrast agent concentration, scanning direction, spatial resolution, joint spacing, and tube current, with known measurements as the reference standard. Materials and Methods: A phantom with nine chambers was constructed. Each chamber had a nylon cylinder encased by sleeves of aluminum and polycarbonate to simulate trabecular bone, cortical bone, and cartilage. Varying simulated cartilage thicknesses and 10 joint space widths were assessed. On 3 days, the phantom was scanned with and without contrast agent administration and with the chamber axes both perpendicular and parallel to the scanner axis. Images were reconstructed at 1.0- and 0.5-mm intervals. Contrast agent concentration and tube current were varied. The simulated cartilage thickness was determined by using image segmentation. Root m...