Cartilage Thickness: Factors Influencing Multidetector CT Measurements in a Phantom Study
- 1 January 2008
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 246 (1) , 133-141
Abstract
Purpose: To prospectively assess in a phantom the reconstruction errors and detection limits of cartilage thickness measurements obtained with multidetector computed tomographic (CT) arthrography, as a function of contrast agent concentration, scanning direction, spatial resolution, joint spacing, and tube current, with known measurements as the reference standard. Materials and Methods: A phantom with nine chambers was constructed. Each chamber had a nylon cylinder encased by sleeves of aluminum and polycarbonate to simulate trabecular bone, cortical bone, and cartilage. Varying simulated cartilage thicknesses and 10 joint space widths were assessed. On 3 days, the phantom was scanned with and without contrast agent administration and with the chamber axes both perpendicular and parallel to the scanner axis. Images were reconstructed at 1.0- and 0.5-mm intervals. Contrast agent concentration and tube current were varied. The simulated cartilage thickness was determined by using image segmentation. Root m...Keywords
This publication has 39 references indexed in Scilit:
- Fat-Suppressed 3D Spoiled Gradient-Echo MRI and MDCT Arthrography of Articular Cartilage in Patients with Hip DysplasiaAmerican Journal of Roentgenology, 2005
- Comparison of multislice CT arthrography and MR arthrography for the detection of articular cartilage lesions of the elbowEuropean Radiology, 2005
- Cartilage Thickness in Cadaveric Ankles: Measurement with Double-Contrast Multi–Detector Row CT Arthrography versus MR ImagingRadiology, 2004
- Cartilage lesions in the ankle joint: comparison of MR arthrography and CT arthrographySkeletal Radiology, 2003
- Assessment of Knee Cartilage in Cadavers with Dual-Detector Spiral CT Arthrography and MR ImagingRadiology, 2002
- Effect of gradient and section orientation on quantitative analysis of knee joint cartilageJournal of Magnetic Resonance Imaging, 2000
- Evaluation of patellar cartilage surface lesions: comparison of CT arthrography and fat-suppressed FLASH 3D MR imagingEuropean Radiology, 1998
- In vivo reproducibility of three-dimensional cartilage volume and thickness measurements with MR imaging.American Journal of Roentgenology, 1998
- Assessment of normal patellar cartilage volume and thickness using MRI: an analysis of currently available pulse sequencesSkeletal Radiology, 1996
- Accuracy of fat-suppressed three-dimensional spoiled gradient-echo FLASH MR imaging in the detection of patellofemoral articular cartilage abnormalities.Radiology, 1996