Combining pedicle screw stimulation with spinal navigation, a protocol to maximize the safety of neural elements and minimize radiation exposure in thoracolumbar spine instrumentation
- 29 April 2015
- journal article
- research article
- Published by Springer Nature in European Spine Journal
- Vol. 25 (6) , 1724-1728
Abstract
Purpose The O-arm-based navigation increases the accuracy of pedicle screw positioning and offers the possibility of performing a 3D scan before wound closure. However, repeating the 3D scan exposes the patient to additional radiation. We combined O-arm navigation with pedicle screw (PS) stimulation followed by a 3D scan to evaluate their accuracy and aimed for the creation of a protocol that maximizes the safety and minimizes radiation. Methods Patients had pedicle screws insertion using O-arm spinal navigation, then had PS triggered electromyography (EMG), and finally a 3D scan to evaluate the accuracy of screw position. Results 447 screws were inserted in 71 patients. In 10 patients, 11 screws needed repositioning. Comparing results of PS triggered EMG responses to the 3D scan, we found: (a) negative stimulation response with negative 3D scan findings, corresponding to 432 acceptable screw position (96.6 %) in 58 patients (81.7 %). In these cases, the redo 3D scan could be avoided. (b) Positive stimulation response with positive 3D scan findings, corresponding to 7 unacceptable screw position (1.5 %) in 6 patients (8.4 %). In these cases, PS stimulation detected malpositioned screws that would be missed without a redo 3D scan. Conclusion We propose a protocol of routinely performing PS stimulation after screw insertion using spinal navigation. In case of positive stimulation, a 3D scan must be performed to rule out a probable screw mal position (6 patients 8.4 %). However, in case of negative stimulation, redo 3D scan can be avoided in 81.7 % of patients.Keywords
This publication has 25 references indexed in Scilit:
- Comparison of Navigated Versus Non-Navigated Pedicle Screw Placement in 260 Patients and 1434 ScrewsJournal of Spinal Disorders & Techniques, 2015
- Routine spinal navigation for thoraco–lumbar pedicle screw insertion using the O-arm three-dimensional imaging system improves placement accuracyJournal of Clinical Neuroscience, 2014
- The accuracy of pedicle screw placement using intraoperative image guidance systemsJournal of Neurosurgery: Spine, 2014
- Fijación lumbar posterior empleando el sistema de imagen quirúrgica O-arm: experiencia inicialNeurocirugia, 2013
- A Prospective Multicenter Registry on the Accuracy of Pedicle Screw Placement in the Thoracic, Lumbar, and Sacral Levels With the Use of the O-arm Imaging System and StealthStation NavigationSpine, 2012
- Radiation Exposure to the Spine Surgeon in Lumbar and Thoracolumbar Fusions With the Use of an Intraoperative Computed Tomographic 3-Dimensional Imaging SystemSpine, 2012
- Electromyographic thresholds after thoracic screw stimulation depend on the distance of the screw from the spinal cord and not on pedicle cortex integrityThe Spine Journal, 2012
- Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniquesEuropean Spine Journal, 2011
- Role of intraoperative neurophysiologic monitoring in lumbosacral spine fusion and instrumentation: a retrospective studyWorld Neurosurgery, 2010
- Lumbar pedicle: surgical anatomic evaluation and relationshipsEuropean Spine Journal, 2001