Dosimetric Characteristics of a PIN Diode for Radiotherapy Application
Authors:
Open Access
- 1 August 2014
- journal article
- research article
- Published by SAGE Publications in Technology in Cancer Research & Treatment
- Vol. 13 (4) , 361-367
Abstract
The PIN diode developed by Bhabha Atomic Research Centre (BARC) was modified for its use as a dosimeter in radiation therapy. For this purpose the diode was mounted on a printed circuit board (PCB) and provided with necessary connections so that its response against irradiation can be recorded by a standard radiotherapy electrometer. The dosimetric characteristics of the diode were studied in Co-60 gamma rays as well as high energy X-rays. The measured sensitivity of this PIN diode is 4 nC/cGy which is about ten times higher than some commercial diode dosimeters. The leakage current from the diode is 0.04 nA. The response of the PIN diode is linear in the range of 20–1000 cGy which covers the full range of radiation dose encountered in radiotherapy treatments. The non-linearity of the diode response is 3.5% at 20 cGy and it is less than 1.5% at higher dose values. Its repeatability is within 0.5%. The angular response variation is about 5.6% within ± 60° with respect to normal beam incidence. The response of the PIN diode at 6 and 18 MV X-rays varies within 2% with respect to its response at Co-60 gamma rays. The source to surface distance (SSD) dependence of the PIN diode was studied for Co-60 beam. It was found that the response of the diode decreases almost linearly relative to given dose for beams with constant collimator setting but increasing SSD (decreasing dose-rate). Within this study the diode response varied by about 2.5% between the maximum and minimum SSD. The dose-rate dependence of the PIN diode for 6 and 15 MV-rays was studied. The variation in response of diode for both energies in the studied dose range is less than 1%. The field size dependence of the PIN diode response is within 1% with respect to the response of ionisation chamber. These studies indicate that the characteristics of the PIN diode are suitable for use in radiotherapy dosimetry.Keywords
This publication has 18 references indexed in Scilit:
- Accelerator beam data commissioning equipment and procedures: Report of the TG‐106 of the Therapy Physics Committee of the AAPMMedical Physics, 2008
- State of the art of in vivo dosimetryRadiation Protection Dosimetry, 2008
- Modeling the instantaneous dose rate dependence of radiation diode detectorsMedical Physics, 2003
- Electronic dosimeters based on solid state detectorsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2001
- Application of P-I-N Diodes and Mosfets for Dosimetry in Gamma and Neutron Radiation FieldsRadiation Protection Dosimetry, 1999
- In vivo dosimetry during conformal therapy of prostatic cancerRadiotherapy and Oncology, 1993
- Personal Neutron Diode DosemeterRadiation Protection Dosimetry, 1993
- A comparison of semiconductor and thermoluminescent dosemeters for in vivo dosimetryBritish Journal of Radiology, 1991
- Dose evaluation to patients irradiated by 60Co beams, by means of direct measurement on the incident and on the exit surfacesRadiotherapy and Oncology, 1985
- Evaluation of the Clinical Use of TLDActa Radiologica: Therapy, Physics, Biology, 1976