Dosimetric accuracy of proton therapy for chordoma patients with titanium implants
- 25 June 2013
- journal article
- Published by Wiley in Medical Physics
- Vol. 40 (7) , 071727
Abstract
Purpose:: To investigate dosimetric errors in proton therapy treatment planning due to titanium implants, and to determine how these affect postoperative passively scattered proton therapy for chordoma patients with orthopedic hardware.Methods:: The presence of titanium hardware near the tumor may affect the dosimetric accuracy of proton therapy. Artifacts in the computed tomography (CT) scan can cause errors in the proton stopping powers used for dose calculation, which are derived from CT numbers. Also, clinical dose calculation algorithms may not accurately simulate proton beam transport through the implants, which have very different properties as compared to human tissue. The authors first evaluated the impact of these two main issues. Dose errors introduced by metal artifacts were studied using phantoms with and without titanium inserts, and patient scans on which a metal artifact reduction method was applied. Pencil‐beam dose calculations were compared to models of nuclear interactions in titanium and Monte Carlo simulations. Then, to assess the overall impact on treatment plans for chordoma, the authors compared the original clinical treatment plans to recalculated dose distributions employing both metal artifact reduction and Monte Carlo methods.Results:: Dose recalculations of clinical proton fields showed that metal artifacts cause range errors up to 6 mm distal to regions affected by CT artifacts. Monte Carlo simulations revealed dose differences >10% in the high‐dose area, and range differences up to 10 mm. Since these errors are mostly local in nature, the large number of fields limits the impact on target coverage in the chordoma treatment plans to a small decrease of dose homogeneity.Conclusions:: In the presence of titanium implants, CT metal artifacts and the approximations of pencil‐beam dose calculations cause considerable errors in proton dose calculation. The spatial distribution of the errors however limits the overall impact on passively scattered proton therapy for chordoma.Keywords
Funding Information
- Massachusetts General Hospital (C06-CA059267)
- Massachusetts General Hospital (C06-CA059267)
This publication has 30 references indexed in Scilit:
- Comprehensive analysis of proton range uncertainties related to patient stopping-power-ratio estimation using the stoichiometric calibrationPhysics in Medicine & Biology, 2012
- CT metal artifact reduction method correcting for beam hardening and missing projectionsPhysics in Medicine & Biology, 2012
- Spot-Scanning-Based Proton Therapy for Extracranial ChordomaInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Phase II Study of High-Dose Photon/Proton Radiotherapy in the Management of Spine SarcomasInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Can megavoltage computed tomography reduce proton range uncertainties in treatment plans for patients with large metal implants?Physics in Medicine & Biology, 2008
- PET/CT imaging for treatment verification after proton therapy: A study with plastic phantoms and metallic implantsMedical Physics, 2007
- The influence of metal artefacts on the range of ion beamsPhysics in Medicine & Biology, 2007
- Treatment Planning for Conformal Proton Radiation TherapyTechnology in Cancer Research & Treatment, 2003
- A pencil beam algorithm for proton dose calculationsPhysics in Medicine & Biology, 1996
- Compensating for heterogeneities in proton radiation therapyPhysics in Medicine & Biology, 1984