Calibration Method for Standard Scattering Plate Calibration System Used in Calibrating Visibility Meter

Authors:
  • 张健张健ZHANGJian
    张健ZHANGJian
  • 张国张国玉ZHANGGuo-yu
    张国玉ZHANGGuo-yu
  • 孙高孙高飞SUNGao-fei
    孙高飞SUNGao-fei
  • 苏拾苏拾SUShi
    苏拾SUShi
  • 张建张建良ZHANGJian-liang
    张建良ZHANGJian-liang

Abstract
In order to realized the rapid and accurate calibrating of the standard scattering plate uesd in calibrating visibility meter, calibrateing the simulated atmospheric visibility by measuring the scattering coefficient of multiple scattering angle of the standard scattering plate, and the calibration system was calibrated. According to the composition and working principle of the calibration system, the calibration method was determined, the calibration chain of the calibration system was established, and the calibration method for each step of the the calibration chain was designed. By calibrating the standard scattering plate with a given scattering coefficient, the veracity of the calibration result of the calibration system and the correctness of the calibration method were verified. The experimental results prove that the error of the calibration results of scattering coefficient of standard scattering plate is 7.93%, the error of the simulating visibility of the standard scatting plate calibrated by the calibration system is 8.61%, satisfied the requirements of the standard scattering plate used in calibrating visibility meter.