Design of an Ankle Rehabilitation Device Using Compliant Mechanisms
Authors:
- 6 January 2011
- journal article
- Published by ASME International in Journal of Medical Devices
- Vol. 5 (1) , 011001
Abstract
In this paper, we present the design, analysis, and testing of an ankle rehabilitation device (ARD), the purpose of which is to improve the efficacy of ankle joint complex (AJC) injury diagnosis and treatment. The ARD enables physicians to quantitatively measure the severity of an injury. This is done by measuring deficiencies in the joint’s range of motion, as well as force, torque, and power output. Evaluation of the relative degree of recovery over time can also reduce the error associated with current methodologies for rehabilitation, which rely on measurements based on the patient’s verbal response. A Wheatstone bridge circuit is used for the measurement of the various parameters as applied to the blades of complementary rotational flexures; the device is designed to measure motion about three axes of rotation in the ankle joint: pitch, roll, and yaw. A full bridge circuit is applied to each axis of rotation, and the use of multiple axes increases anatomically accurate measurement, enabling characterization of coupled motions. The device has flexibility and a range of motion such that it can be adjusted to take measurements of multiple different degrees of plantar or dorsiflexion of the AJC. The ARD is able to measure both range of motion, force, and torque output simultaneously. Experimental results show that there is significant coupled motion among the ankle joint rotations but that it is highly dependent on a subject’s own physical development.Keywords
This publication has 11 references indexed in Scilit:
- Ankle Rehabilitation via Compliant MechanismsJournal of Medical Devices, 2010
- Treatment of Common Deficits Associated with Chronic Ankle InstabilitySports Medicine, 2009
- Optimal design of a compliant mechanism with circular notch flexure hingesProceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science, 2007
- Benefits of exercise for community-dwelling older adults 11No commercial party having a direct financial interest in the results of the research supporting this article has or will confer a benefit upon the authors(s) or upon any organization with which the author(s) is/are associated.Archives of Physical Medicine and Rehabilitation, 2004
- Ankle and Subtalar Kinematics Measured with Intracortical Pins during the Stance Phase of WalkingFoot & Ankle International, 2004
- A Comparison of 3 Hand-Held Dynamometers Used to Measure Hip Abduction StrengthJournal of Strength & Conditioning Research, 2003
- Muscle Power of the Ankle Flexors Predicts Functional Performance in Community‐Dwelling Older WomenJournal of the American Geriatrics Society, 2001
- A Stewart Platform-Based System for Ankle TelerehabilitationAutonomous Robots, 2001
- Anatomy and Biomechanics of the First RayPhysical Therapy, 1999
- Miniature load cells for the measurement of foot-ground reaction forces and centre of foot pressure during gaitJournal of Biomedical Engineering, 1986