Fabrication of the Enzyme‐less Voltammetric Bilirubin Sensor Based on Sol‐gel Imprinted Polymer
- 30 October 2019
- journal article
- research article
- Published by Wiley in Electroanalysis
- Vol. 32 (3) , 479-488
Abstract
Measuring the serum bilirubin (BR) level is part of the common test for newborns, since high concentrations of bilirubin in the blood can cause brain damage and even death for them. Therefore, achieving a fast, accurate and cost effective method for bilirubin assay in medical diagnosis and Medical management of patients with jaundice, especially neonates, is very important. In this study, an electrochemical sensor was developed to detect low amounts of bilirubin in serum samples. The basis of the work is the BR imprinted polymer synthesis and the use of this polymer as a recognition element in the production of the bilirubin voltammetric sensor. For this purpose, bilirubin imprinted polymer was synthesized using the simple sol gel method and the prepared polymer is characterized by SEM and FT‐IR. In the next step, this polymer was applied to modify the carbon paste electrode and provide a sensitive and selective sensor for the BR molecule. The proposed sensor was able to measure the bilirubin in the concentration range of 1 to 100 μM after optimizing the operating conditions. The limit of detection (LOD) and limit of quantification (LOQ) was calculated 0.75 μM and 2.5 μM respectively. It is also applicable to bilirubin assay in serum samples.Keywords
Funding Information
- Center for International Scientific Studies and Collaboration
- Ministry of Science Research and Technology, Iran
This publication has 46 references indexed in Scilit:
- Spectrofluorimetric determination of bilirubin in serum samplesLuminescence, 2011
- Promethazine determination in plasma samples by using carbon paste electrode modified with molecularly imprinted polymer (MIP): Coupling of extraction, preconcentration and electrochemical determinationElectrochimica Acta, 2010
- Via zinc(II) protoporphyrin to the synthesis of poly(ZnPP-MAA-EGDMA) for the imprinting and selective binding of bilirubinBiomaterials, 2009
- Highly sensitive method for quantitative determination of bilirubin in biological fluids and tissuesJournal of Chromatography B, 2008
- Electrocatalytic oxidation of bilirubin at ferrocenecarboxamide modified MWCNT–gold nanocomposite electrodesMicrochimica Acta, 2008
- Molecularly imprinted polymer using β-cyclodextrin as functional monomer for the efficient recognition of bilirubinAnalytica Chimica Acta, 2008
- A portable potentiostat for the bilirubin-specific sensor prepared from molecular imprintingBiosensors and Bioelectronics, 2007
- Amperometric detection of bilirubin from a micro-sensing electrode with a synthetic bilirubin imprinted poly(MAA-co-EGDMA) filmBiosensors and Bioelectronics, 2006
- Fluorescence reaction of bilirubin with zinc ion in dimethyl sulfoxide and its application to assay of total bilirubin in serumAnalytica Chimica Acta, 1998
- Enzyme electrode for the determination of bilirubinElectroanalysis, 1996