Accuracy of corneal power measurements by a new Scheimpflug camera combined with Placido-disk corneal topography for intraocular lens power calculation in unoperated eyes
Authors:
- Giacomo SaviniGSGiacomo Savini
- Piero BarboniPBPiero Barboni
- Michele CarbonelliMCMichele Carbonelli
- Kenneth J. HofferKHKenneth J. Hoffer
- 1 May 2012
- journal article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 38 (5) , 787-792
Abstract
To assess the accuracy of the corneal power measurements with a new Scheimpflug camera combined with Placido-disk corneal topography (Sirius) (combined Scheimpflug camera-topographer) for intraocular lens (IOL) power calculation in unoperated eyes and compare the results with those by a validated corneal topographer (Keratron) (validated topographer). Private practice. Case series. Consecutive patients having phacoemulsification and in-the-bag IOL implantation were studied. Intraocular lens power was calculated using the Hoffer Q, Holladay 1, and SRK/T formulas; the axial length, as measured by ultrasound immersion biometry; and 3 corneal power measurements: validated topographer simulated keratometry (K); combined Scheimpflug camera-topographer simulated K (derived from anterior corneal curvature only); combined Scheimpflug camera-topographer mean pupil power (derived from anterior and posterior corneal curvatures through ray tracing). The prediction error was calculated as the difference between the predicted refraction and the refraction measured 1 month postoperatively. When the corneal power measurements from the combined Scheimpflug camera-topographer were used, the mean absolute error (MAE) ranged between 0.23 diopter (D) ± 0.24 (SD) (simulated K and Hoffer Q formula) and 0.33 ± 0.23 D (mean pupil power and SRK/T formula). There were no statistically significant differences between the MAE generated by the simulated Ks of the 2 devices with any of the 3 formulas. Both corneal power measurements (simulated K and mean pupil power) provided by the new combined Scheimpflug camera-topographer were successfully entered into third-generation IOL power calculation formulas in unoperated eyes. Dr. Hoffer is the author of the Hoffer Q formula and owns the EyeLab, which sells Hoffer Programs. No other author has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 18 references indexed in Scilit:
- Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topographyJournal of Cataract & Refractive Surgery, 2011
- Accuracy of a dual Scheimpflug analyzer and a corneal topography system for intraocular lens power calculation in unoperated eyesJournal of Cataract & Refractive Surgery, 2011
- Repeatability and Reproducibility of Corneal Curvature Measurements Using the Pentacam and Keratron Topography SystemsJournal of Refractive Surgery, 2009
- Scheimpflug photography keratometry readings for routine intraocular lens power calculationJournal of Cataract & Refractive Surgery, 2009
- Prediction of the effective postoperative (intraocular lens) anterior chamber depthJournal of Cataract & Refractive Surgery, 2006
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Corneal Geometry Reconstruction with the Keratron VideokeratographerOptometry and Vision Science, 1997
- Prospective evaluation of early visual and refractive effects with small clear corneal incision for cataract surgeryJournal of Cataract & Refractive Surgery, 1996
- Assessment of Radial Aspheres by the Arc-Step Algorithm as Implemented by the Keratron KeratoscopeAmerican Journal of Ophthalmology, 1995
- A three-part system for refining intraocular lens power calculationsJournal of Cataract & Refractive Surgery, 1988