A multi-channel soft X-ray flat-field spectrometer
Authors:
- 1 January 1998
- conference paper
- Published by AIP Publishing in AIP Conference Proceedings
Abstract
The development of a three channel soft X-ray spectrometer which was specifically designed to resolve the angular intensity distribution of laser generated high harmonics is reported. The spectrometer uses a novel design of three tapered glancing angle mirrors to subtend angles of and in the horizontal plane from the source. These angles were selected to characterise emission within an f5 angular cone, assuming radial symmetry. The soft X-ray mirrors are tapered in width to form an elliptical profile under the application of a bending moment. Design studies demonstrated that by the application of different bending moments, a similar shaped tapered mirror can be used for imaging in each of the three channels. The total subtended solid angle of the instrument in the three channels is 1 millisteradian. The measured resolution of the system is reported as well as the absolute throughput efficiencies in the spectral range 4–40 nm.