Abstract
Purpose: To validate a new 2D-3D registration method of fusion imaging during aortic repair in a system prepared only for 3D-3D registration and to compare radiation doses and accuracy. Materials and Methods: The study involved 189 patients, including 94 patients (median age 70 years; 85 men) who underwent abdominal endovascular aneurysm repair (EVAR) with 2D-3D fusion on an Artis zee imaging system and 95 EVAR patients (median age 70 years; 81 men) from a prior study who had 3D-3D registration done using cone beam computed tomography (CBCT). For the 2D-3D registration, an offline CBCT of the empty operating table was imported into the intraoperative dataset and superimposed on the preoperative computed tomography angiogram (CTA). Then 2 intraoperative single-frame 2D images of the skeleton were aligned with the patient’s skeleton on the preoperative CTA to complete the registration process. A digital subtraction angiogram was done to correct any misalignment of the aortic CTA volume. Values are given as the median [interquartile range (IQR) Q1, Q3]. Results: The 2D-3D registration had an accuracy of 4.0 mm (IQR 3.0, 5.0) after bone matching compared with the final correction with DSA (78% within 5 mm). By applying the 2D-3D protocol the radiation exposure (dose area product) from the registration of the fusion image was significantly reduced compared with the 3D-3D registration [1.12 Gy∙cm2 (IQR 0.41, 2.14) vs 43.4 Gy∙cm2 (IQR 37.1, 49.0), respectively; p<0.001). Conclusion: The new 2D-3D registration protocol based on 2 single-frame images avoids an intraoperative CBCT and can be used for fusion imaging registration in a system originally designed for 3D-3D only. This 2D-3D registration protocol is accurate and leads to a significant reduction in radiation exposure.Keywords
Funding Information
- Skånes universitetssjukhus
- Emil and Wera Cornell foundation
- Svenska Sällskapet för Medicinsk Forskning
- Governmental funding of clinical research within the national health services
- Hulda Almroth Foundation
- Swedish Heart and Lung Foundation
This publication has 11 references indexed in Scilit:
- Radiation-Induced DNA Damage in Operators Performing Endovascular Aortic RepairCirculation, 2017
- Endovascular versus open repair of abdominal aortic aneurysm in 15-years' follow-up of the UK endovascular aneurysm repair trial 1 (EVAR trial 1): a randomised controlled trialThe Lancet, 2016
- Fusion Imaging to Support Endovascular Aneurysm Repair Using 3D-3D RegistrationJournal of Endovascular Therapy, 2016
- Comparison of Radiation Exposure Associated With Intraoperative Cone-Beam Computed Tomography and Follow-up Multidetector Computed Tomography Angiography for Evaluating Endovascular Aneurysm RepairsJournal of Endovascular Therapy, 2016
- JOURNAL CLUB: Cancer Risks in U.S. Radiologic Technologists Working With Fluoroscopically Guided Interventional Procedures, 1994-2008American Journal of Roentgenology, 2016
- The effects of combining fusion imaging, low-frequency pulsed fluoroscopy, and low-concentration contrast agent during endovascular aneurysm repairJournal of Vascular Surgery, 2016
- Three-dimensional fusion computed tomography decreases radiation exposure, procedure time, and contrast use during fenestrated endovascular aortic repairJournal of Vascular Surgery, 2015
- Impact of Hybrid Rooms with Image Fusion on Radiation Exposure during Endovascular Aortic RepairEuropean Journal of Vascular and Endovascular Surgery, 2014
- Evaluation of Visceral Artery Displacement by Endograft Delivery System InsertionJournal of Endovascular Therapy, 2014
- Intraoperative C-arm cone-beam computed tomography in fenestrated/branched aortic endograftingJournal of Vascular Surgery, 2011