Clinical evaluation of ocular biometry of dual Scheimpflug analyzer, GALILEI G6 and swept source optical coherence tomography, ANTERION
Authors:
Open Access
- 4 March 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-8
Abstract
To evaluate the performance of a new swept source optical coherence tomography optical biometer, ANTERION, in ocular biometry and intraocular lens (IOL) calculation compared with the reference standard of Dual Scheimpflug Analyzer (GALILEI, G6). A prospective comparative study was conducted in a tertiary eye center. Cataract patients were scanned with both devices in a random fashion, and parameters from the devices were analyzed in terms of mean difference and intraclass correlation coefficient (ICC). Bland–Altman plots were performed to compare agreement between the devices. Ninety-six eyes from 96 patients were enrolled for evaluation. With the exception of ACD, all parameters were significantly different, but excellent agreement was revealed for all of them. The mean difference in axial length was 0.03 mm, and ICC was 0.999. Calculated IOL power with Barrett formula revealed that 93.75% were within 1 diopter and the prediction error was 0.03 diopter. Biometry of the devices were arithmetically different. However, the mean difference of the key factors in IOL calculation were small and appeared to be negligible for the purposes of clinical application. The performance of ANTERION was comparable to that of G6 in biometric measurement and IOL calculation; however, the devices cannot be used interchangeably.Keywords
Funding Information
- Rajavithi research (241/2563, 241/2563, 241/2563, 241/2563, 241/2563, 241/2563)
This publication has 23 references indexed in Scilit:
- Comparison of normal and keratoconic corneas by Galilei Dual-Scheimpflug AnalyzerContact Lens and Anterior Eye, 2013
- Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topographyJournal of Cataract & Refractive Surgery, 2011
- Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyesJournal of Cataract & Refractive Surgery, 2011
- Accuracy of a dual Scheimpflug analyzer and a corneal topography system for intraocular lens power calculation in unoperated eyesJournal of Cataract & Refractive Surgery, 2011
- Effects of Pterygium on Ocular AberrationsCornea, 2010
- Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographerJournal of Cataract & Refractive Surgery, 2010
- Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetryIndian Journal Of Ophthalmology, 2010
- Biometry and intraocular lens power calculationCurrent Opinion in Ophthalmology, 2008
- Calculation of intraocular lens power: a reviewActa Ophthalmologica Scandinavica, 2007
- Calculation of intraocular lens power: a reviewActa Ophthalmologica Scandinavica, 2007