Comparison of dynamic and static hot wire anemometer calibrations for velocity perturbation measurements

Authors:

Abstract
The sensitivity of constant temperature hot wire anemometers to velocity perturbations in an air flow was measured by the traditional static calibration technique as well as by the recently revived dynamic method in which the wire is shaken in a steady flow. Contrary to previously published findings, the two methods gave calibration constants which agreed to within an experimental standard error of 2% in the velocity range of 3-32 m s-1. A new signal processing technique was used for the dynamic calibration method which gave better accuracy in the presence of electronic noise and wind tunnel turbulence than the RMS method used to date.

This publication has 4 references indexed in Scilit: