Imaging Artifacts of Medical Instruments in Ultrasound-Guided Interventions
Open Access
- 1 October 2007
- journal article
- research article
- Published by Wiley in Journal of Ultrasound in Medicine
- Vol. 26 (10) , 1303-1322
Abstract
Objectives Real‐time 3‐dimensional (3D) ultrasound imaging has the potential to become a dominant imaging technique for minimally invasive surgery. One barrier to its widespread use is that surgical instruments generate imaging artifacts, which can obfuscate their location, orientation, and geometry and obscure nearby tissue. The purpose of this study was to identify and describe the types of artifacts which could be produced by metallic instruments during interventions guided by 3D ultrasound imaging. Methods Three imaging studies were performed. First, imaging artifacts from stainless steel rods were identified in vitro and acoustically characterized. Second, 3 typical minimally invasive instruments were imaged (in vitro and in vivo), and their artifacts were analyzed. The third study compared the intensity of imaging artifacts (in vitro and in vivo) from stainless steel rods with rods composed of 3 different materials and stainless steel rods with roughened and coated surfaces. Results For the stainless steel rods, all observed artifacts are described and illustrated, and their physical origins are explained. Artifacts from the 3 minimally invasive instruments are characterized with the use of the artifacts observed with the rods. Finally, it is shown that artifacts can be greatly reduced through the use of alternate materials or by surface modification. Conclusions Instrument artifacts in 3D ultrasound images can be more confusing than those from the same instruments imaged in 2 dimensions. Real‐time 3D ultrasound imaging can, however, be used effectively for in vivo imaging of minimally invasive instruments by using artifact mitigation techniques, including careful selection of probe and incision locations, as well as by instrument modification.Keywords
Funding Information
- National Institutes of Health National Institute of Biomedical Imaging and Bioengineering (RO1 EB003052)
- National Institutes of Health Bioengineering Research Partnership (R01 HL073647)
This publication has 48 references indexed in Scilit:
- The aurora sign: an ultrasonographic sign suggesting parenchymal lung diseaseBritish Journal of Radiology, 2003
- Sources and impact of artifacts on clinical three‐dimensional ultrasound imagingUltrasound in Obstetrics & Gynecology, 2000
- Sonographically guided percutaneous biopsy of small (3 cm or less) massesAmerican Journal of Roentgenology, 1988
- Detection of Foreign Bodies in Soft Tissues: Experimental Comparison of Ultrasonography and Xeroradiography1988
- The sonographic appearance and contrast enhancement of puncture needlesJournal of Clinical Ultrasound, 1983
- The use of real‐time ultrasound to enhance fetoscopic visualizationJournal of Clinical Ultrasound, 1983
- Slice‐thickness artifacts in gray‐scale ultrasoundJournal of Clinical Ultrasound, 1981
- Ultrasonic appearance of metallic foreign bodies in parenchymal organsJournal of Clinical Ultrasound, 1981
- Range ambiguities in real‐time ultrasoundJournal of Clinical Ultrasound, 1981
- Sound Scattering by Solid Cylinders and SpheresThe Journal of the Acoustical Society of America, 1951