Is the Sequence of SuperAlarm Triggers More Predictive Than Sequence of the Currently Utilized Patient Monitor Alarms?
- 30 June 2016
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 64 (5) , 1023-1032
Abstract
Objective: Our previous studies have shown that “code blue” events can be predicted by SuperAlarm patterns that are multivariate combinations of monitor alarms and laboratory test results cooccurring frequently preceding the events but rarely among control patients. Deploying these patterns to the monitor data streams can generate SuperAlarm sequences. The objective of this study is to test the hypothesis that SuperAlarm sequences may contain more predictive sequential patterns than monitor alarms sequences. Methods: Monitor alarms and laboratory test results are extracted from a total of 254 adult coded and 2213 control patients. The training dataset is composed of subsequences that are sampled from complete sequences and then further represented as fixeddimensional vectors by the term frequency inverse document frequency method. The information gain technique and weighted support vector machine are adopted to select the most relevant features and train a classifier to differentiate sequences between coded patients and control patients. Performances are assessed based on an independent dataset using three metrics: sensitivity of lead time (Sen L @T), alarm frequency reduction rate (AFRR), and work-up to detection ratio (WDR). Results:The performance of 12-h-long sequences of SuperAlarm can yield a Sen L @2 of 93.33%, an AFRR of 87.28%, and a WDR of 3.01. At an AFRR = 87.28%, Sen L @2 for raw alarm sequences and discretized alarm sequences are 73.33% and 70.19%, respectively. At a WDR = 3.01, Sen L @2 are 49.88% and 43.33%. Conclusion and Significance: The results demonstrate that SuperAlarm sequences indeed outperform monitor alarm sequences and suggest that one can focus on sequential patterns from SuperAlarm sequences to develop more precise patient monitoring solutions.Keywords
Funding Information
- UCSF Middle-Career Scientist Award (R01NHLBI128679, R18H5022860)
- UCSF Institute of Computational Health Sciences
This publication has 33 references indexed in Scilit:
- Reducing False Alarms of Intensive Care Online‐Monitoring Systems: An Evaluation of Two Signal Extraction AlgorithmsComputational and Mathematical Methods in Medicine, 2011
- Medical device alarmsBiomedical Engineering / Biomedizinische Technik, 2011
- Intensive care unit alarms—How many do we need?*Critical Care Medicine, 2010
- Monitor Alarm Fatigue: Standardizing Use of Physiological Monitors and Decreasing Nuisance AlarmsAmerican Journal of Critical Care, 2010
- Performance of feature-selection methods in the classification of high-dimension dataPattern Recognition, 2009
- Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveformJournal of Biomedical Informatics, 2008
- A discretization algorithm based on Class-Attribute Contingency CoefficientInformation Sciences, 2008
- Alarm Algorithms in Critical Care MonitoringAnesthesia & Analgesia, 2006
- The use of the area under the ROC curve in the evaluation of machine learning algorithmsPattern Recognition, 1997