A comparison of early redo surgery rates in Mosaic porcine and Perimount bovine pericardial valves
- 20 March 2018
- journal article
- research article
- Published by Oxford University Press (OUP) in European Journal of Cardio-Thoracic Surgery
- Vol. 54 (4) , 724-728
Abstract
OBJECTIVES: The objective of this study was to compare rates of redo surgery for the Medtronic Mosaic 305 A Porcine Prosthesis and the Carpentier-Edwards Perimount Pericardial Aortic Bioprosthesis 2900. METHODS: This was a single-centre retrospective observational study. We included all 1018 patients who underwent aortic valve replacement with a Mosaic (n = 216) or Perimount (n = 809) bioprosthesis between October 2000 and August 2008. The total follow-up was 1508 patient-years for the Mosaic valve and 5813 for the Perimount valve. The maximal follow-up and interquartile range were 14.8 and 7.0 years for the Mosaic valve and 15.1 and 5.6 years for the Perimount valve, respectively. A propensity score-weighted version of the Cox model, Kaplan-Meier analysis and multivariate regression model was used. RESULTS: Despite no statistical difference in the number of non-structural valve deterioration cases between valves, redo surgery occurred earlier in 10 (4.6%) Mosaic than for 17 (2.1%) Perimount valves (P = 0.02) and was required for structural valve deterioration in 5 (2.3%) Mosaic valves when compared with 7 (0.9%; P = 0.04) Perimount valves. Four of 5 Mosaic failures occurred before 5 years, whereas all Perimount failures occurred after 5 years. Redo surgery for non-structural valve deterioration occurred in 3 patients with Mosaic valves (1.4%) and no patients with Perimount valves. Surgery for the remaining patients with Perimount valves was due to infection or aortic disease. CONCLUSIONS: Early redo surgery for structural valve degeneration was uncommon but occurred earlier for the Mosaic porcine than the Perimount bovine pericardial replacement aortic valve.Keywords
Funding Information
- King's College London
This publication has 14 references indexed in Scilit:
- A comparison of outcomes between bovine pericardial and porcine valves in 38 040 patients in England and Wales over 10 yearsEuropean Journal of Cardio-Thoracic Surgery, 2014
- Long-term results of the Medtronic Mosaic porcine bioprosthesis in the aortic positionThe Journal of Thoracic and Cardiovascular Surgery, 2014
- Late Survival After Aortic Valve Replacement With the Perimount Versus the Mosaic BioprosthesisThe Annals of Thoracic Surgery, 2014
- Randomized Trial of Carpentier-Edwards Supraannular Prosthesis Versus Mosaic Aortic Prosthesis: 6 Year ResultsThe Annals of Thoracic Surgery, 2013
- Hemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of a prospectively randomized studyEuropean Journal of Cardio-Thoracic Surgery, 2011
- Choice of Prosthetic Heart Valve in AdultsJACC, 2010
- Early stenosis of Medtronic Mosaic porcine valves in the aortic positionThe Journal of Thoracic and Cardiovascular Surgery, 2009
- Bovine pericardial versus porcine stented replacement aortic valves: Early results of a randomized comparison of the Perimount and the Mosaic valvesThe Journal of Thoracic and Cardiovascular Surgery, 2008
- Guidelines for reporting mortality and morbidity after cardiac valve interventionsThe Journal of Thoracic and Cardiovascular Surgery, 2008
- Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten YearsThe Annals of Thoracic Surgery, 2007