Repeatability of Corneal Densitometry Measurements Using a Scheimpflug Camera in Healthy Normal Corneas
Authors:
Open Access
- 1 January 2022
- journal article
- research article
- Published by Medknow in Journal of Current Ophthalmology
- Vol. 34 (1) , 50-55
Abstract
Purpose: To determine the repeatability of corneal densitometry measured by the Scheimpflug imaging system. Methods: This cross-sectional study was conducted on photorefractive keratectomy candidates. One eye of each participant underwent imaging using Pentacam HR three times, 10 min apart. The repeatability of densitometry measurements was evaluated in four concentric annuli around the corneal apex and in different corneal depths. The repeatability of the measurements was evaluated using the intraclass correlation coefficient (ICC), repeatability coefficient (RC), and coefficient of variation (CV). The difference of repeatability between layers and zones was tested by tolerance index (TI). Results: Sixty eyes of sixty patients with a mean age of 27.76 ± 3.93 years were studied. Half of the participants were female (n = 30, 50%). ICC was above 0.9 in all corneal parts. The posterior layer and central zones showed the least variability of densitometry measurements considering the CV values. The RC was 2.06, 1.17, and 0.92 in anterior, central, and posterior layers, respectively. The RC was 0.88, 0.71, 1.51, and 4.56 in 0–2, 2–6, 6–10, and 10–12 mm circles, respectively. Only the reliability of densitometry in 10–12 mm annulus was statistically lower than the central zone (TI = 0.71). Conclusions: Corneal densitometry measurements provided by the Pentacam had good repeatability. The repeatability of densitometry measurements decreased from the center to the periphery (with an exception for 0–2 mm and 2–6 mm) and from the posterior to the anterior of the cornea. The reliability of the 10–12 mm zone was markedly less than other zones.Keywords
This publication has 25 references indexed in Scilit:
- Reliability of Pentacam HR Thickness Maps of the Entire Cornea in Normal, Post–Laser In Situ Keratomileusis, and Keratoconus EyesAmerican Journal of Ophthalmology, 2016
- Corneal structure and transparencyProgress in Retinal and Eye Research, 2015
- Tolerance and Relative Utility: Two Proposed Indices for Comparing Change in Clinical Measurement Noise Between Different Populations (Repeatability) or Measurement Methods (Agreement)Investigative Ophthalmology & Visual Science, 2015
- Effect of keratoconus grades on repeatability of keratometry readings: Comparison of 5 devicesJournal of Cataract & Refractive Surgery, 2015
- Normative Values for Corneal Densitometry Analysis by Scheimpflug Optical AssessmentInvestigative Ophthalmology & Visual Science, 2014
- Corneal Densitometry as an Indicator of Corneal HealthOphthalmology, 2012
- Natural history of corneal haze after collagen crosslinking for keratoconus and corneal ectasia: Scheimpflug and biomicroscopic analysisJournal of Cataract & Refractive Surgery, 2010
- Repeatability and validity of lens densitometry measured with Scheimpflug imagingJournal of Cataract & Refractive Surgery, 2009
- Intraobserver and interobserver repeatability of curvature and aberrometric measurements of the posterior corneal surface in normal eyes using Scheimpflug photographyJournal of Cataract & Refractive Surgery, 2009
- Wound Healing After Excimer Laser Keratomileusis (Photorefractive Keratectomy) in MonkeysArchives of Ophthalmology (1950), 1990