The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants
Open Access
- 8 June 2021
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 92 (5) , 621-627
Abstract
Background and purpose — We noted several adverse events in patients in whom the first version of the STRYDE limb-lengthening nail (NuVasive Specialized Orthopaedics, San Diego, CA) had been implanted. Pain, osteolysis, peri- osteal reactions, and cortical hypertrophy at the nail junc- tion were noted. Here, we present the analysis of 23 retrieved STRYDE implants. Materials and methods — We undertook visual inspec- tion of the retrieved nails and screws, mechanical evaluation of the junction, micro-CT analyses, microscopic inspection of the bushing, screws, screw holes, and separated parts of the implants. Positive material identification (PMI) and energy-dispersive X-ray spectroscopy (EDS) were used to analyze the chemical composition. The hardness of the mate- rial was also investigated. Results — 20/23 retrieved nails had visible signs of cor- rosion, i.e., discoloration at the telescopic junction. Micro- CT verified corrosion attacks in 12/12 scanned bushings. Corrosion, predominantly mechanically assisted crevice corrosion, was observed at the locking screws and screw holes in 20/23 nails. Biological material inside the nail was observed in addition to oozing from the junction of 2 nails during hardware removal, which was experimentally repro- ducible. Notably, the mechanical construction of the bushing changed from PRECICE P2 to STRYDE nails. Interpretation — STRYDE nails are not hermetically sealed, and liquid can pass the bushing. Biodur 108 itself is corrosion resistant; however, mechanically assisted crev- ice corrosion of the bushing, locking screws, and screw holes may be aggravated due to manufacturing aiming for increased strength and hardness of the alloy.Keywords
This publication has 33 references indexed in Scilit:
- Magnetically Controlled Growing Rod for Early-Onset Scoliosis: Systematic Review and Meta-AnalysisWorld Neurosurgery, 2019
- Limb lengthening and deformity correction with externally controlled motorized intramedullary nails: evaluation of 50 consecutive lengtheningsActa Orthopaedica, 2018
- A retrieval analysis of the Precice intramedullary limb lengthening systemBone & Joint Research, 2018
- A Comparison of the Device-Related Complications of Intramedullary Lengthening Nails Using a New Classification SystemBioMed Research International, 2017
- Eliminating detrimental effect of cold working on pitting corrosion resistance in high nitrogen austenitic stainless steelsCorrosion Science, 2017
- Use of internal lengthening nails in post-traumatic sequelaeInternational Orthopaedics, 2017
- Manganese Effects on Repassivation Kinetics and SCC Susceptibility of High Mn–N Austenitic Stainless Steel AlloysJournal of the Electrochemical Society, 2007
- Effect of nitrogen on crevice corrosion in austenitic stainless steelCorrosion Science, 2006
- On the pitting corrosion resistance of nitrogen alloyed cold worked austenitic stainless steelsCorrosion Science, 2002
- The influences of microstructure and nitrogen alloying on pitting corrosion of type 316L and 20 wt.% Mn-substituted type 316L stainless steelsCorrosion Science, 2001