Evaluation of a Veterinary Glucometer for Use in Horses
Authors:
- 1 May 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Veterinary Internal Medicine
- Vol. 24 (3) , 617-621
Abstract
Background: Glucose assessment and regulation are important factors in the treatment of hospitalized horses and foals. Hypothesis/Objectives: The purpose of this study was to compare glucose measurement by a veterinary glucometer, adjusted by code for use in horses and foals, to a reference chemistry analyzer. It was hypothesized that the veterinary glucometer and reference analyzer would yield similar results and that interpretation of glucose values obtained from a veterinary glucometer would result in clinically appropriate decisions. Animals: Fifty blood samples from adult horses and 50 blood samples from neonatal foals admitted to the Colorado State University Veterinary Hospital or Equine Reproduction Laboratory for evaluation. Methods: Glucose concentrations from fresh whole blood samples were evaluated in duplicate with a veterinary glucometer and these values were compared with those obtained with a reference plasma chemistry analyzer. The accuracy of glucometer measurement was evaluated with a Clarke error grid. Results: The veterinary glucometer accurately measured whole blood glucose concentrations in both horses and foals when compared with a reference plasma chemistry analyzer. Nearly 97% of the glucometer values obtained in this study would have resulted in appropriate clinical decisions based on the Clarke error grid analysis. Conclusions and Clinical Importance: The veterinary glucometer evaluated has potential utility for point-of-care whole blood glucose evaluation in both horses and foals.Keywords
This publication has 22 references indexed in Scilit:
- Assessing the Performance of Handheld Glucose Testing for Critical CareDiabetes Technology & Therapeutics, 2008
- Blood Glucose Concentrations in Critically Ill Neonatal FoalsJournal of Veterinary Internal Medicine, 2008
- Comparison of the Accu-Chek Aviva Point-of-Care Glucometer with Blood Gas and Laboratory Methods of Analysis of Glucose Measurement in Equine Emergency PatientsJournal of Veterinary Internal Medicine, 2008
- Evaluation of point-of-care glucose testing accuracy using locally-smoothed median absolute difference curvesClinica Chimica Acta, 2008
- Blood Glucose in Horses with Acute Abdominal DiseaseJournal of Veterinary Internal Medicine, 2007
- Agreement between point‐of‐care glucometry, blood gas and laboratory‐based measurement of glucose in an equine neonatal intensive care unitJournal of Veterinary Emergency and Critical Care, 2007
- Challenges of implementing Point-of-Care Testing (POCT) glucose meters in a pediatric acute care settingClinical Biochemistry, 2004
- Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose AnalyzerAmerican Journal of Clinical Pathology, 2000
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- The distribution of sodium, potassium and glucose in the blood of some mammalsComparative Biochemistry and Physiology, 1967