Robotic‐arm assisted total knee arthroplasty has a learning curve of 16 cases and increased operative time of 12 min
Open Access
- 12 August 2022
- journal article
- research article
- Published by Wiley in Anz Journal of Surgery
- Vol. 92 (11) , 2974-2979
Abstract
Background: Robotic‐arm assisted systems are increasingly used for knee arthroplasty, however introduction of new systems can involve a learning curve. We aimed to define the learning curve in terms of operative time and component placement/sizing of a robotic system for total knee arthroplasty (TKA) in a team of experienced surgeons, and to investigate mid‐term patient outcomes.Methods: A total of 101 consecutive patients underwent primary robotic‐arm assisted TKA by three surgeons (mean 2 year follow‐up). Operative times, component placement, implant sizing and reoperations were recorded. Cumulative Summation (CUSUM) was used to analyse learning curves. Patient outcomes were compared between learning and proficiency phases.Results: The learning curve was 16 cases, with a 12‐min increase in operative time (P < 0.01). Once proficiency was achieved, the greatest time reductions were seen for navigation registration (P = 0.003) and bone preparation (P < 0.0001). A learning curve was found with polyethylene (PE) insert sizing (P = 0.01). No differences were found between learning and proficiency groups in terms of implant survival (100% and 97%, respectively, NS) or patient‐reported outcome measures at 2 years (NS).Conclusion: Introduction of a robotic‐arm assisted system for TKA led to increased operative times for navigation registration and bone preparation, and a learning curve with PE insert sizing. No difference in patient outcomes between learning and proficiency groups at 2 years was found. These findings can inform surgeons' expectations when starting to use robotic‐assisted systems.Keywords
This publication has 31 references indexed in Scilit:
- The minimal clinically important difference in the Oxford knee score and Short Form 12 score after total knee arthroplastyKnee Surgery, Sports Traumatology, Arthroscopy, 2013
- The “Forgotten Joint” as the Ultimate Goal in Joint ArthroplastyThe Journal of Arthroplasty, 2012
- Polyethylene thickness is a risk factor for wear necessitating insert exchangeInternational Orthopaedics, 2011
- A Complication-based Learning Curve From 200 Reverse Shoulder ArthroplastiesClinical Orthopaedics and Related Research, 2011
- Development and preliminary testing of the new five-level version of EQ-5D (EQ-5D-5L)Quality of Life Research, 2011
- “Thicker” Polyethylene Bearings Are Associated With Higher Failure Rates in Primary Total Knee ArthroplastyThe Journal of Arthroplasty, 2010
- Learning Curve With Minimally Invasive Unicompartmental Knee ArthroplastyThe Journal of Arthroplasty, 2010
- The use of the Oxford hip and knee scoresThe Journal of Bone and Joint Surgery. British volume, 2007
- Polyethylene in Knee ArthroplastyClinical Orthopaedics and Related Research, 2005
- Number of Polyethylene Particles and Osteolysis in Total Joing Replacements: A Quanatitive Study Using a Tissue-Digestion MethodThe Journal of Bone and Joint Surgery. British volume, 1997