Gradient-induced voltages on 12-lead ECGs during high duty-cycle MRI sequences and a method for their removal considering linear and concomitant gradient terms
- 23 June 2015
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 75 (5) , 2204-2216
Abstract
Purpose To restore 12‐lead electrocardiographic (ECG) signal fidelity inside MRI by removing magnetic field gradient‐induced voltages during high gradient duty cycle sequences. Theory and Methods A theoretical equation was derived to provide first‐ and second‐order electrical fields induced at individual ECG electrodes as a function of gradient fields. Experiments were performed at 3T on healthy volunteers using a customized acquisition system that captured the full amplitude and frequency response of ECGs, or a commercial recording system. The 19 equation coefficients were derived via linear regression of data from accelerated sequences and were used to compute induced voltages in real‐time during full resolution sequences to remove ECG artifacts. Restored traces were evaluated relative to ones acquired without imaging. Results Measured induced voltages were 0.7 V peak‐to‐peak during balanced steady state free precession (bSSFP) with the heart at the isocenter. Applying the equation during gradient echo sequencing, three‐dimensional fast spin echo, and multislice bSSFP imaging restored nonsaturated traces and second‐order concomitant terms showed larger contributions in electrodes further from the magnet isocenter. Equation coefficients are evaluated with high repeatability (ρ = 0.996) and are dependent on subject, sequence, and slice orientation. Conclusion Close agreement between theoretical and measured gradient‐induced voltages allowed for real‐time removal. Prospective estimation of sequence periods in which large induced voltages occur may allow hardware removal of these signals. Magn Reson Med 75:2204–2216, 2016.Keywords
Funding Information
- National Institutes of Health (R03-EB013873-01A1, P41EB015898)
- American Heart Association (10SDG261039)
This publication has 24 references indexed in Scilit:
- 3DQRS: A method to obtain reliable QRS complex detection within high field MRI using 12‐lead electrocardiogram tracesMagnetic Resonance in Medicine, 2014
- A 1.5T MRI‐conditional 12‐lead electrocardiogram for MRI and intra‐MR interventionMagnetic Resonance in Medicine, 2013
- Adaptive noise cancellation to suppress electrocardiography artifacts during real‐time interventional MRIJournal of Magnetic Resonance Imaging, 2011
- Nonlinear Bayesian Filtering for Denoising of Electrocardiograms Acquired in a Magnetic Resonance EnvironmentIEEE Transactions on Biomedical Engineering, 2010
- Independent component analysis-based artefact reduction: application to the electrocardiogram for improved magnetic resonance imaging triggeringPhysiological Measurement, 2009
- Measurement of electric fields due to time-varying magnetic field gradients using dipole probesPhysics in Medicine & Biology, 2007
- Noise Cancellation Signal Processing Method and Computer System for Improved Real-Time Electrocardiogram Artifact Correction During MRI Data AcquisitionIEEE Transactions on Biomedical Engineering, 2007
- Improving automatic analysis of the electrocardiogram acquired during magnetic resonance imaging using magnetic field gradient artefact suppressionJournal of Electrocardiology, 2006
- Suppression of MR gradient artefacts on electrophysiological signals based on an adaptive real-time filter with LMS coefficient updatesMagnetic Resonance Materials in Physics, Biology and Medicine, 2005
- Analytic calculations of the E-fields induced by time-varying magnetic fields generated by cylindrical gradient coilsMagnetic Resonance in Medicine, 2000