Brewster Angle Apparatus for Thin Film Index Measurements

Authors:

Abstract
For refractive index measurements of solid thin films when film absorption is negligible, the Brewster angle method described by Abeles [ Compt. rend. 228, 553 ( 1949)] is extremely convenient. An apparatus is described which facilitates routine index measurements using this method. A corner mirror system elimates the need for manually tracking the reflected beam as the angle of incidence is varied. Settings are reproducible to within about 0.1% of the index value under optimun conditions, the results agreeing with those of other observers.

This publication has 1 reference indexed in Scilit: