Current and electric charge pulse shapes obtained for the collection of only one electron-hole pair released by an ionizing event in the sensitive volume of a coaxial Ge(Li) detector cooled at liquid nitrogen temperature, were calculated. By considering the pairs generated in the close vioinity of the detector electrodes, as well as those created in an intermediate region, maximum and minimum durations of the current pulses have been found. Low electric field intensity is assumed (charge carrier velocity directly proportional to the electric field intensity).Keywords
This publication has 5 references indexed in Scilit:
- Mobilite des electrons et des trous a 77°K dans le germanium compense au lithiumNuclear Instruments and Methods, 1967
- Pulse formation and transit time of charge carriers in semiconductor junction detectorsNuclear Instruments and Methods, 1964
- On the induced charge in semiconductor detectorsNuclear Instruments and Methods, 1963
- On the Induced Current and Energy Balance in ElectronicsProceedings of the IRE, 1941
- Currents Induced by Electron MotionProceedings of the IRE, 1939