Fiber dispersion in time domain measurements compromising the accuracy of determination of optical properties of strongly scattering media
- 1 January 2003
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Biomedical Optics
- Vol. 8 (3) , 512-516
Abstract
Limitations in the accuracy of measurements of optical properties (absorption and scattering coefficients) of strongly scattering media that are due to temporal dispersion inside the detection fibers of a time-of-flight setup were investigated for a tissue-like phantom. It is shown that the absorption and reduced scattering coefficients may be overestimated by up to 90% (depending on length and numerical aperture of the fibers) if the instrumental response measurements for the setup are performed by direct illumination of the tip of the detection fibers with collimated short laser pulses. However, the accuracy can be improved significantly if a thin layer of scattering media is used in front of the detection fibers during the response measurements. The relevance of the investigated dispersion effects is discussed with respect to frequency-domain measurements as well. © 2003 Society of Photo-Optical Instrumentation Engineers.Keywords
This publication has 17 references indexed in Scilit:
- Optical properties of human skin in the near infrared wavelength range of 1000 to 2200 nmJournal of Biomedical Optics, 2001
- Breast imaging technology: Probing physiology and molecular function using optical imaging - applications to breast cancerBreast Cancer Research, 2000
- A 32-channel time-resolved instrument for medical optical tomographyReview of Scientific Instruments, 2000
- Compact tissue oximeter based on dual-wavelength multichannel time-resolved reflectanceApplied Optics, 1999
- A solid tissue phantom for photon migration studiesPhysics in Medicine & Biology, 1997
- Shades of gray matter: Noninvasive optical images of human brain reponses during visual stimulationPsychophysiology, 1995
- The influence of glucose concentration upon the transport of light in tissue-simulating phantomsPhysics in Medicine & Biology, 1995
- A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivoMedical Physics, 1992
- Quantitation of time- and frequency-resolved optical spectra for the determination of tissue oxygenationAnalytical Biochemistry, 1991
- Effect of photosensitizer concentration in tissue on the penetration depth of photoactivating lightLasers in Medical Science, 1986