Skin-interfaced microfluidic devices with one-opening chambers and hydrophobic valves for sweat collection and analysis
- 18 June 2020
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 20 (15) , 2635-2645
Abstract
Soft, skin-interfaced microfluidic platforms are capable of capturing, storing, and assessing sweat chemistry and total sweat loss, which provides essential insight into human physiological health. However, sweat loss from the outlet of the microfluidic devices often leads to deviation of the measured concentration of the biomarker or electrolyte from the actual value. Here, we introduce hydrophobic valves at the junction of the chamber and the microfluidic channel as a new chamber design to reduce sweat evaporation. Because the advancing front of the liquid in the hydrophilic microchannel is blocked by the hydrophobic valve, the fluid flows into the chambers, forms the initial meniscus, and completely fills the chambers along the initial meniscus. Fluid dynamic modeling and numerical simulations provide critical insights into the sweat sampling mechanism into the chambers. With significantly reduced evaporation and contamination, the sweat sample can be easily stored for a long time for later analysis when in situ analysis is limited. Additionally, the design with multiple chambers can allow sequential generation of sweat collection at different times for long-term analysis. The in situ real-time measurements of the sweat loss and pH value analysis from the human subject demonstrate the practical utility of the devices in collecting, storing, and analyzing the sweat generated from sweat glands on the skin.Keywords
Funding Information
- National Natural Science Foundation of China (11872326, 11702236)
- Education Department of Hunan Province (18B089)
- Pennsylvania State University
- Natural Science Foundation of Hunan Province (2018JJ2379, 2018JJ2396)
- Tsinghua University (2019KF1102)
- American Chemical Society Petroleum Research Fund (59021-DNI7)
This publication has 33 references indexed in Scilit:
- Assembling Metal–Organic Frameworks into the Fractal Scale for Sweat SensingACS Applied Materials & Interfaces, 2019
- Passive sweat collection and colorimetric analysis of biomarkers relevant to kidney disorders using a soft microfluidic systemLab on a Chip, 2019
- Exercise intensity effects on total sweat electrolyte losses and regional vs. whole-body sweat [Na+], [Cl−], and [K+]European Journal of Applied Physiology, 2018
- Soft, Skin‐Interfaced Microfluidic Systems with Wireless, Battery‐Free Electronics for Digital, Real‐Time Tracking of Sweat Loss and Electrolyte CompositionSmall, 2018
- Body map of regional vs. whole body sweating rate and sweat electrolyte concentrations in men and women during moderate exercise-heat stressJournal of Applied Physiology, 2018
- Stretchable temperature-sensing circuits with strain suppression based on carbon nanotube transistorsNature Electronics, 2018
- A soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweatScience Translational Medicine, 2016
- The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implicationsBiomicrofluidics, 2015
- Concept and development of an autonomous wearable micro-fluidic platform for real time pH sweat analysisSensors and Actuators B: Chemical, 2012
- The pathopysiology of heat stroke: an integrative view of the final common pathwayScandinavian Journal of Medicine & Science in Sports, 2011