Silicon Microfabricated Column with Microfabricated Differential Mobility Spectrometer for GC Analysis of Volatile Organic Compounds
Authors:
- 25 October 2005
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 77 (23) , 7563-7571
Abstract
A 3.0-m-long, 150-μm-wide, 240-μm-deep channel etched in a 3.2-cm-square silicon chip, covered with a Pyrex wafer, and coated with a dimethyl polysiloxane stationary phase is used for the GC separation of volatile organic compounds. The column, which generates ∼5500 theoretical plates, is temperature-programmed in a conventional convection oven. The column is connected through a heated transfer line to a microfabricated differential mobility spectrometer. The spectrometer incorporates a 63Ni source for atmospheric-pressure chemical ionization of the analytes. Nitrogen or air transport gas (flow 300 cm3/min) drives the analyte ions through the cell. The spectrometer operates with an asymmetric radio frequency (RF) electric field between a pair of electrodes in the detector cell. During each radio frequency cycle, the ion mobility alternates between a high-field and a low-field value (differential mobility). Ions oscillate between the electrodes, and only ions with an appropriate differential mobility reach a pair of biased collectors at the downstream end of the cell. A compensation voltage applied to one of the RF electrodes is scanned to allow ions with different differential mobilities to pass through the cell without being annihilated at the RF electrodes. A unique feature of the device is that both positive and negative ions are detected from a single experiment. The combined microfabricated column and detector is evaluated for the analysis of volatile organic compounds with a variety of functionalities.Keywords
This publication has 25 references indexed in Scilit:
- Stop-Flow Programmable Selectivity with a Dual-Column Ensemble of Microfabricated Etched Silicon Columns and Air as Carrier GasAnalytical Chemistry, 2005
- Separation of Ions from Explosives in Differential Mobility Spectrometry by Vapor-Modified Drift GasAnalytical Chemistry, 2004
- Design, Fabrication, and Evaluation of Microfabricated Columns for Gas ChromatographyAnalytical Chemistry, 2004
- Differential mobility spectrometry of chlorocarbons with a micro-fabricated drift tubeThe Analyst, 2004
- Development of a preconcentration unit for a SAW sensor micro array and its use for indoor air quality monitoringSensors and Actuators B: Chemical, 2003
- Parylene gas chromatographic column for rapid thermal cyclingJournal of Microelectromechanical Systems, 2002
- Fast GC instrumentation and analysis for field applicationsField Analytical Chemistry & Technology, 2001
- Thermal analysis and simulation of the microchannel flow in miniature thermal conductivity detectorsSensors and Actuators A: Physical, 2000
- High-Speed, Vacuum-Outlet GC Using Atmospheric-Pressure Air as Carrier GasAnalytical Chemistry, 1999
- Review and summary of a silicon micromachined gas chromatography systemIEEE Transactions on Components, Packaging, and Manufacturing Technology: Part B, 1998