Alteration of the Pressure Setting of a Codman-Hakim Programmable Valve by a Television-Case Report-
Authors:
Open Access
- 1 January 2006
- journal article
- case report
- Published by Japan Neurosurgical Society in Neurologia medico-chirurgica
- Vol. 46 (8) , 405-407
Abstract
A 7-year-old girl presented in semicomatose condition. She had received a ventriculoperitoneal shunt using a Codman-Hakim programmable valve. Head computed tomography demonstrated hydrocephalus and head radiography showed that the pressure setting of the shunt valve had changed to 60 mmH2O from 40 mmH2O. The pressure setting was returned to 40 mmH2O, and she was discharged because her clinical symptoms and hydrocephalus improved. One month later, she lost consciousness again and was transported to our hospital. Hydrocephalus and shunt valve pressure of 50 mmH2O were noted, and the pressure setting was returned to 40 mmH2O again. She was discharged without complications. We suspected that the valve pressure was caused by close contact with a television, because the patient tended to touch a television with her head during play. The valve pressure did not change after the television was placed on a high stand out of reach. We should recognize that there are many sources of weak magnetic fields that may influence a programmable valve in everyday life.Keywords
This publication has 9 references indexed in Scilit:
- Effect of Vagus Nerve Stimulator Magnet on Programmable Shunt SettingsNeurosurgery, 2004
- Clinical experience with the use of a shunt with an adjustable valve in children with hydrocephalusJournal of Neurosurgery, 2003
- Electromagnetic field hazards involving adjustable shunt valves in hydrocephalusJournal of Neurosurgery, 2002
- Pressure changes observed in Codman-Medos programmable valves following magnetic exposure and fillipingChild's Nervous System, 2001
- Seven years of clinical experience with the programmable Codman Hakim valve: a retrospective study of 583 patientsJournal of Neurosurgery, 2000
- Spontaneous resetting of a CodmanMedos programmable shuntBritish Journal Of Neurosurgery, 2000
- A Randomized, Controlled Study of a Programmable Shunt Valve versus a Conventional Valve for Patients with HydrocephalusNeurosurgery, 1999
- Transcutaneous Pressure-adjustable Valves and Magnetic Resonance Imaging: An Ex Vivo Examination of the Codman-Medos Programmable Valve and the Sophy Adjustable Pressure ValveNeurosurgery, 1997
- Safety of the adjustable pressure ventricular valve in magnetic resonance imaging: problems and solutionsNeuroradiology, 1992