Count rate performances of TTVO3: the CEA-LETI time-of-flight positron emission tomograph
Authors:
- 1 January 1991
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 10 (3) , 261-266
Abstract
The authors present the count rate performance of the CEA-LETI TTVO3 time-of-flight positron emission tomography (PET) system using both physical measurements and H(2)(15)O bolus human myocardial studies. They also present a comparison between the counting statistics of H(2)(15)O brain studies performed on this machine and on the latest available high-resolution brain bismuth germanate (BGO) tomograph, the ECAT 953B/31. During the 80 mCi cerebral blood flow study, the count rate reached 100 K events/s, and the same experiment performed on a high-resolution BGO brain machine gave only a 30% increase in signal. These results demonstrate that TTVO3 is particularly suitable for H(2)(15)O flow studies.Keywords
This publication has 11 references indexed in Scilit:
- Physical characteristics of the ECAT 953B/31: a new high resolution brain positron tomographIEEE Transactions on Medical Imaging, 1991
- Physical characteristics of TTV03, a new high spatial resolution time-of-flight positron tomographIEEE Transactions on Nuclear Science, 1990
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- Monte Carlo simulation of BaF/sub 2/ detectors used in time-of-flight positron emission tomographyIEEE Transactions on Nuclear Science, 1990
- EEC concerted action on cellular degeneration and regeneration studied with PETEuropean Journal of Nuclear Medicine and Molecular Imaging, 1990
- Noninvasive quantitation of myocardial blood flow in human subjects with oxygen-15-labeled water and positron emission tomographyJACC, 1989
- Performance measurements of the SP3000/UW time-of-flight positron emission tomographIEEE Transactions on Nuclear Science, 1988
- The Effect of Live Time in Components of a Positron Tomograph on Image QuantificationIEEE Transactions on Nuclear Science, 1987
- Performance of Positron Imaging Systems as a Function of Energy Threshold and Shielding DepthIEEE Transactions on Medical Imaging, 1984
- Correction for Scattered Radiation in a Ring Detector Positron Camera by Integral Transformation of the ProjectionsJournal of Computer Assisted Tomography, 1983