Fabrication and Characterization of p-type Inverted Coaxial Point Contact (ICPC) Detectors with a-Ge Dual-Blocking Contacts
Authors:
- S. A. PanamaldeniyaSPS. A. Panamaldeniya
- K. M. DongKDK. M. Dong
- D. M. MeiDMD. M. Mei
Preprint
Abstract
We report the fabrication and characterization of two p-type inverted coaxial point contact (ICPC) high-purity germanium (HPGe) detectors, SAP16 and SAP17, produced from USD-grown crystals with net impurity concentrations of $\sim 3\times10^{10}\,\mathrm{cm^{-3}}$. Both devices employ \emph{thin} amorphous-germanium (a-Ge) dual-blocking contacts, implemented here for the first time on ICPC detectors, to provide bipolar charge blocking while limiting dead-layer thickness. Electrical tests at 76~K demonstrate stable operation with picoampere-level leakage currents and sub-pF capacitance: SAP17 reached $\sim 4.62$~pA at the maximum tested bias (500~V) and operated stably at 400~V with $C\simeq 0.503$~pF. \emph{Meanwhile,} SAP16 achieved superior spectroscopic performance, with energy resolutions of 2.42\% at 59.5~keV and 0.36\% at 662~keV. Gamma-ray spectroscopy with $^{241}$Am and $^{137}$Cs shows that modest geometric differences lead to measurable changes in depletion behavior and charge-collection uniformity, consistent with electrostatic modeling. Angular-response measurements further reveal pronounced directional sensitivity at 59.5~keV, whereas the 662~keV response is essentially isotropic over the measured range. These results validate thin a-Ge dual-blocking contacts for ICPC HPGe detectors and highlight geometry-driven trade-offs among leakage current, depletion, and energy resolution relevant to low-background and low-threshold applications.
Keywords
All Related Versions
This publication has 23 references indexed in Scilit:
- Pulse shape discrimination performance of inverted coaxial Ge detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018
- Ultra-low noise mechanically cooled germanium detectorNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Crystal growth and detector performance of large size High-purity Ge crystalsMaterials Science in Semiconductor Processing, 2015
- Inter-electrode charge collection in high-purity germanium detectors with amorphous semiconductor contactsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015
- Leakage current in high-purity germanium detectors with amorphous semiconductor contactsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015
- Determination of dead-layer variation in HPGe detectorsApplied Radiation and Isotopes, 2014
- Dislocation density control in high-purity germanium crystal growthJournal of Crystal Growth, 2014
- A novel HPGe detector for gamma-ray tracking and imagingNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- A Pulse Shape Analysis technique for the MAJORANA experimentNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- Amorphous-semiconductor-contact germanium-based detectors for gamma-ray imaging and spectroscopyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007