Abstract
A titanium dioxide (TiO2) nanoparticles (NP)-deposited column by the liquid phase deposition (LPD) method was exploited for open tubular capillary electrochromatographic (OTCEC) separation of proteins and peptides. With the phosphate buffer, the direction of electroosmotic flow (EOF) for the TiO2 NP-deposited capillary was the same as that for the bare fused silica capillary, but the magnitude of EOF was much smaller at pH≧7. Its OTCEC performance was then evaluated with four proteins (bovine serum albumin, ovalbumin, hemoglobin, and myoglobin) by varying parameters of the background electrolyte (BGE), such as pH value, composition, and concentration. Good separation was achieved with the phosphate buffer (40 mM, pH 9.0) and an applied voltage of 15 kV. In comparison with the bare fused silica capillary, the relatively high resolving power of the TiO2 NP-deposited capillary would mainly result from the ligand exchange interaction between analytes and the phosphate adsorbed on the TiO2. Finally, separation of the proteolytic digest mixture was also performed with the TiO2 NP-deposited capillary at well defined conditions, indicating its potential applicability for biological analysis.

This publication has 39 references indexed in Scilit: