We have developed an instrument to visually monitor a material cooled in cryogenics. A temperature controlled video scope works well in a helium gas environment and in the vicinity of a surface of liquid helium. The instrument enables us to monitor the situation of a material located at low temperature. Utilizing the instrument, we have succeeded in recovering electrical contact to the electrical socket of a superconducting magnet covered with contaminant of solid air. The socket was dug out by a drill with monitoring by the instrument. Behaviors of solid air in cryogenic instruments for an injection of a helium gas were also visually observed.